a D:GROUP company
00020EN Rev.09 – 12/2012
11
5-6 LEFT/RIGHT ARROWS
Use the arrow keys to scroll the menu: if pressed the following data will appear on the
display:
U.V. Lamp Residual Lifetime:
Shows the operating time of the U.V.lamp pre-set
by the user with the appropriate keys. The LCD will display (for example) "U.V.
TIME=XXXX h" . When such time is over, the message "U.V. LIFETIME OVER"
will appear on the line below.
1, 2, 3
Residual lifetime of filters
: it is the operation time of the filters installed
in
the
cabinet
that
can
be
programmed
by
the
user.
The LCD will display (for example)" RES. TIME FILTER 1=XXXX:XX h:min".
When such time is over, the message "CHECK FILTER (i.e.) 1". will appear on
the line below.
The filters installed in the cabinet follow the numbering listed below
TYPE of FILTER
NUMBER
MAIN HEPA
1
EXHAUST HEPA
2
ACTIVE CARBON
3
LAF Power: it is shown indirectly by the power supply voltage of the
motor, expressed as percentage of max. load voltage displayed also in
proportion by a bar.
The display shows the notice(es.): "MOT.LAF = XX % " (max.100%).
Operating Time:
Shows the operating time of the cabinet from the moment
when the main switch is positioned on "I"
The LCD will display (for example) "WORK TIME=XXXXXh”. This value cannot
be reset.
7 LIGHT
This switches on the fluorescent light; when enabled, the display shows "Light
on". Switching on the fluorescent light automatically the U.V. lamp switches off.
8 SOCKET
This
supplies
voltage; when enabled, if the PCB for the automatic regulation is
installed the display shows "POWER ON". The global current for all the sockets
installed on the cabinet is 4 Amps.
9 GAS (optional)
This activates the control for opening/closing the gas electrovalve; when enabled, if the PCB for
the automatic regulation is installed the display shows "GAS ON". It operates
only when the ventilation is running to prevent possible over-heating and risks of
damaging the HEPA filter.
10 ESC
ESC key deletes the operation of data input and goes back to the starting
condition
.
When an alarm condition occurs, which is shown also by the message appearing
on the LCD. By pushing "ESC" the alarm stops sounding. If the cause of the
alarm is not resolved after 2 min the buzzer starts to sound again.
Summary of Contents for FlowFAST H
Page 32: ...a D GROUP company 00020EN Rev 09 12 2012 32...
Page 33: ...a D GROUP company 00020EN Rev 09 12 2012 33...
Page 35: ...a D GROUP company 00020EN Rev 09 12 2012 35 14 B SIDE DIAGRAM...
Page 36: ...a D GROUP company 00020EN Rev 09 12 2012 36 15 CABINET MOUNTING...
Page 37: ...a D GROUP company 00020EN Rev 09 12 2012 37...
Page 38: ...a D GROUP company 00020EN Rev 09 12 2012 38 16 ASSEMBLY of SUPPORT TABLE...
Page 39: ...a D GROUP company 00020EN Rev 09 12 2012 39 17 ASSEMBLY OF THE CABINET ON THE TABLE optional...
Page 40: ...a D GROUP company 00020EN Rev 09 12 2012 40 18 SIDE DIAGRAM WITH U V LAMP optional...
Page 42: ...a D GROUP company 00020EN Rev 09 12 2012 42 20 WIRING DIAGRAM...
Page 43: ...a D GROUP company 00020EN Rev 09 12 2012 43...
Page 44: ...a D GROUP company 00020EN Rev 09 12 2012 44...