TROUBLESHOOTING PROBABLE CAUSES OF MALFUNCTIONS
PROBLEM
CAUSE
REMEDY
Cabinet does not work
–
the electricity supply has
been cut off at the mains
–
Electronic board out of order
–
Blown fuse(s)
–
check the voltage input to the
cabinet
–
Replace PCB
–
Replace fuses
Alarm: “HEPA filters check”
Main HEPA filter clogged
Replace HEPA filter
HEPA filters clogged .
Replace HEPA filters.
“minimum LAF alarm”.
[Possible contamination of the
product but protection of the
environment]
The main motor-fan does not work
Check the terminal voltage of the
power card of the main motor-fan
Check F7 fuse on the power card
Replace the power card
Replace the microprocessor card
“Black-out ” Alarm
(probable exchange of air between
the work chamber and the outside
and possible contamination of the
environment)
Blackout
Check the feeding cable, the
connection plug/socket, the power
supply line
Press “ESC” [7] key to silence the
alarm
Alarm “Sensors failure”
Failure of sensor XX
Replace XX sensor.
Alarm: “No encoder input LAF”
No signal from LAF flow sensor
Replace the LAF sensor
Summary of Contents for FlowFAST H
Page 32: ...a D GROUP company 00020EN Rev 09 12 2012 32...
Page 33: ...a D GROUP company 00020EN Rev 09 12 2012 33...
Page 35: ...a D GROUP company 00020EN Rev 09 12 2012 35 14 B SIDE DIAGRAM...
Page 36: ...a D GROUP company 00020EN Rev 09 12 2012 36 15 CABINET MOUNTING...
Page 37: ...a D GROUP company 00020EN Rev 09 12 2012 37...
Page 38: ...a D GROUP company 00020EN Rev 09 12 2012 38 16 ASSEMBLY of SUPPORT TABLE...
Page 39: ...a D GROUP company 00020EN Rev 09 12 2012 39 17 ASSEMBLY OF THE CABINET ON THE TABLE optional...
Page 40: ...a D GROUP company 00020EN Rev 09 12 2012 40 18 SIDE DIAGRAM WITH U V LAMP optional...
Page 42: ...a D GROUP company 00020EN Rev 09 12 2012 42 20 WIRING DIAGRAM...
Page 43: ...a D GROUP company 00020EN Rev 09 12 2012 43...
Page 44: ...a D GROUP company 00020EN Rev 09 12 2012 44...