Model 640 & 640CC Instruction Manual
D01206676
Revision A February 2016
43
necessary.
Faulty heater.
Replace.
Faulty solid state relay.
Replace.
Temperature display is
erratic.
Faulty thermocouple.
Replace thermocouple.
Loose connection in
thermocouple wiring.
Check for loose wiring and
correct if necessary.
Temperature displays and
unusually high number
(>1500
°
F)
Open circuit in
thermocouple.
Replace thermocouple.
Open circuit in
thermocouple circuitry
Check thermocouple circuitry
for open circuits or loose
connections. Correct if
necessary.
Oil comes out pump muffler.
Pump may or may not
stroke.
AIR TO CYLINDER
valve has been opened
while cylinder is
pressurized, flooding the
airlines and pump
pneumatic side with oil.
Remove air lines and drain oil.
Remove pump and
disassemble pump pneumatic
side to remove oil. If
contamination is not too
severe, you may let the pump
stroke under no pressure until
the oil is removed. This tends
to create an oil mist that will
fill the entire lab.
Air lubricator is putting
too much oil in the pump
air.
Adjust lubricator to supply
less oil.