Model 640 & 640CC Instruction Manual
D01206676
Revision A February 2016
12
Certain components are supported during shipment with wooden blocks, foam
padding, and plastic ties to prevent damage. Open the front doors and remove all
the packing materials that would interfere with the operation of the instrument
before powering the instrument.
Some components such as touch screen LCD monitors (if applicable) and computer
control modules are be removed from the instrument prior to shipment and may be
in a separate container to prevent damage. These devices must be reinstalled before
operating the instrument. Locate the box containing the touch screen monitors and
remove them from their shipping containers.
While standing at the rear of the instrument, align the four mounting holes in the
black monitor housing with the four holes in the front panel. With the monitor
screen facing the front of the instrument, secure the monitor to the front panel with
the four screws provided. Connect the 15-pin video and USB cables from the
monitor to the back of the control box. You must also connect the power cord from
the monitor to an A/C outlet on the back of the instrument.
If the instrument is equipped with auxiliary standalone temperature controllers,
they must be removed from their shipping containers and installed into the sleeves
in the front panel. Simply slide the controllers into the sleeves and snap them into
place. In order to protect the internal mechanisms of the pressure gauges, the
needles are fixed in place with a screw and a small metal bracket. This bracket must
be removed prior to using the instrument.
On 640 and 640CC consistometers, the counterweight in each swivel arm has been
locked in place prior to shipment to prevent the weight from damaging the swivel
arm. Prior to removing the plug from the pressure vessel, loosen the bolt holding
the counterweight in place until the counterweight is free to move. The control box
should be connected to the instrument by connecting the 5-pin power cord to the
back of the instrument frame as well as the supplied USB cable.
The instrument will not operate without this USB cable connected
from the computer to the machine.
Before operating the instrument, check the bottom of the
consistometer for loose screws or bolts that may have loosened
and/or fallen out during shipment. This is particularly true for
overseas shipments.
There are two power connectors on the back of the electrical box. An
uninterruptible power supply (UPS) may be connected between these power