Working with measurement tools
You can also set an audible alarm (there will be a “beep” when the alarm is triggered).
17.8.4
Procedure
There are different procedures for setting up an alarm for a spot, box, circle, or line and for
a difference calculation.
17.8.4.1
Setting up an alarm for a spot
Follow this procedure:
1. To select the spot, touch the tool on the screen. The tool is now displayed with a frame.
2. Push the joystick—or touch and hold the tool. This displays a context menu.
3. Use the joystick to go to
(
Set alarm on spot
).
4. Push the joystick. This displays a dialog box where you can define the settings for the
alarm.
•
Alarm condition
: The condition that triggers the alarm. Applicable values are
Above
,
Below
, or
Off
.
•
Alarm limit
: The temperature value that will be the critical condition when an alarm is
triggered or not.
•
Alarm sound
: Applicable values are
Beep
or
No sound
.
5. Push the joystick. This closes the dialog box.
17.8.4.2
Setting up an alarm for a box, circle, or line
Note
This procedure assumes that you have previously set the camera to display at least
one value (maximum, minimum, or average) in the result table.
Follow this procedure:
1. To select the measurement tool, touch the tool on the screen. The tool is now dis-
played with one or more handles.
2. Push the joystick—or touch and hold the tool. This displays a context menu.
3. Use the joystick to go to (depending on tool)
,
, or
(
Max/Min/Avg/Alarm
) or
(
Graph/Max/Min/Avg/Alarm
).
4. Push the joystick. This displays a submenu.
5. Select
(
Set alarm
).
6. Push the joystick. This displays a dialog box where you can define the settings for the
alarm.
•
Alarm condition
: The condition that triggers the alarm. Applicable values are
Above
,
Below
, or
Off
.
•
Select measurement
: Applicable settings are the values you have previously de-
fined (
Max
,
Min
, and/or
Avg
).
•
Alarm limit
: The temperature value that will be the critical condition when an alarm is
triggered or not.
•
Alarm sound
: Applicable values are
Beep
or
No sound
.
7. Push the joystick. This closes the dialog box.
17.8.4.3
Setting up an alarm for a difference calculation
Note
• You can set up an alarm for a difference calculation when defining user presets, or
when editing an image in the archive.
• This procedure assumes that you have previously set up a difference calculation.
#T559954; r. AL/37426/37426; en-US
69
Summary of Contents for T10 series
Page 1: ...User s manual FLIR T10xx series ...
Page 2: ......
Page 3: ......
Page 5: ...User s manual FLIR T10xx series T559954 r AL 37426 37426 en US v ...
Page 6: ......
Page 14: ......
Page 217: ...Mechanical drawings 27 T559954 r AL 37426 37426 en US 203 ...