Fakir T10 series User Manual Download Page 1

User’s manual
FLIR T10xx series

Summary of Contents for T10 series

Page 1: ...User s manual FLIR T10xx series ...

Page 2: ......

Page 3: ......

Page 4: ...edést és jogi nyilatkozatot Nota importante Prima di utilizzare il dispositivo è importante leggere capire e seguire tutte le istruzioni avvertenze precauzioni ed esclusioni di responsabilità legali 重要な注意 デバイスをご使用になる前に あらゆる指示 警告 注意事項 および免責条項をお読み頂き その内容を理解して従ってください 중요한 참고 사항 장치를 작동하기 전에 반드시 다음의 사용 설명서와 경고 주의사항 법적 책임제한을 읽고 이해하며 따라야 합니다 Viktig Før du bruker enheten må du lese forstå og følge instruks...

Page 5: ...User s manual FLIR T10xx series T559954 r AL 37426 37426 en US v ...

Page 6: ......

Page 7: ...ntation updates 9 3 7 Important note about this manual 9 3 8 Note about authoritative versions 10 4 Customer help 11 4 1 General 11 4 2 Submitting a question 11 4 3 Downloads 12 5 Introduction 13 5 1 General description 13 5 2 Key benefits 13 6 Quick start guide 14 6 1 Procedure 14 7 About FLIR Tools Tools 15 7 1 Introduction 15 7 2 Workflow 16 7 2 1 General 16 7 2 2 Figure 16 7 2 3 Explanation 16...

Page 8: ...d regulations 27 11 Screen elements 28 11 1 Figure 28 11 2 Explanation 28 11 3 Status icons and indicators 28 11 4 Image overlay information 29 12 Navigating the menu system 30 12 1 General 30 12 2 Navigating using the joystick 30 13 Handling the camera 31 13 1 Charging the battery 31 13 1 1 Using the power supply to charge the battery 31 13 1 2 Using the stand alone battery charger to charge the ...

Page 9: ...p lens 42 13 14 1 General 42 13 14 2 Attaching the close up lens 42 13 14 3 Removing the close up lens 44 13 15 Calibrating the compass 46 13 15 1 Procedure 46 14 Saving and working with images 47 14 1 About image files 47 14 1 1 General 47 14 1 2 File naming convention 47 14 1 3 Image capacity 47 14 1 4 About UltraMax 47 14 2 Saving an image 47 14 2 1 General 47 14 2 2 Procedure 48 14 3 Previewin...

Page 10: ...15 4 1 General 55 15 4 2 Procedure 55 15 5 Changing the color palette 56 15 5 1 General 56 15 5 2 Procedure 57 15 6 Changing the object parameters 57 15 7 Calibrating the camera 58 15 7 1 General 58 15 7 2 Manual calibration 58 15 8 Hiding all overlay 58 15 8 1 General 58 15 8 2 Procedure 59 16 Working with image modes 60 16 1 General 60 16 2 Image examples 60 16 3 Selecting the image mode 61 17 W...

Page 11: ...therms 73 18 3 1 About the Condensation alarm 73 18 3 2 About the Insulation alarm 73 18 3 3 Setting up condensation and insulation alarms 73 19 Annotating images 74 19 1 General 74 19 2 Adding a note 74 19 2 1 General 74 19 2 2 Procedure 74 19 3 Adding a table 74 19 3 1 General 74 19 3 2 Procedure 75 19 4 Adding a voice annotation 75 19 4 1 General 75 19 4 2 Procedure 76 19 5 Adding a sketch 76 1...

Page 12: ...c 28 138 26 12 FLIR T1030sc 45 145 26 13 FLIR T1040 12 152 26 14 FLIR T1040 28 159 26 15 FLIR T1040 45 166 26 16 FLIR T1050sc 12 173 26 17 FLIR T1050sc 28 180 26 18 FLIR T1050sc 45 187 26 19 IR lens f 36 mm 28 with case 194 26 20 Close up lens 3 51 µm with case 196 26 21 IR lens f 21 2 mm 45 with case 197 26 22 IR lens f 83 4 mm 12 with case 199 26 23 FLIR T10xx SC kit 201 27 Mechanical drawings 2...

Page 13: ... 225 34 Glossary 226 35 Thermographic measurement techniques 229 35 1 Introduction 229 35 2 Emissivity 229 35 2 1 Finding the emissivity of a sample 229 35 3 Reflected apparent temperature 233 35 4 Distance 233 35 5 Relative humidity 233 35 6 Other parameters 233 36 History of infrared technology 234 37 Theory of thermography 237 37 1 Introduction 237 37 2 The electromagnetic spectrum 237 37 3 Bla...

Page 14: ......

Page 15: ...f operation Expendable parts are excluded from the warranty In the case of a defect in a product covered by this warranty the product must not be fur ther used in order to prevent additional damage The purchaser shall promptly report any defect to FLIR Systems or this warranty will not apply FLIR Systems will at its option repair or replace any such defective product free of charge if upon inspect...

Page 16: ...nges and improvements on any of the products without prior notice 1 7 Patents One or several of the following patents and or design patents may apply to the products and or features Additional pending patents and or pending design patents may also apply 000279476 0001 000439161 000499579 0001 000653423 000726344 000859020 001106306 0001 001707738 001707746 001707787 001776519 001954074 002021543 0...

Page 17: ...p www gnu org licenses lgpl 2 1 en html Retrieved May 27 2015 1 8 2 Fonts Source Han Sans https github com adobe fonts source han sans blob master LICENSE txt Retrieved May 27 2015 1 8 3 Fonts DejaVu http dejavu fonts org wiki License Retrieved May 27 2015 T559954 r AL 37426 37426 en US 3 ...

Page 18: ...247 issue 1 of Industry Cana da Operation is subject to the following two conditions 1 this device may not cause harmful interference and 2 this device must accept any interference received including interference that may cause undesired operation WARNING This device is granted pursuant to the Japanese Radio Law 電波法 and the Japanese Telecommunica tions Business Law 電気通信事業法 This device should not b...

Page 19: ...n tainers before you use a liquid The liquids can be dangerous Injury to persons can occur CAUTION Do not point the infrared camera with or without the lens cover at strong energy sources for example devices that cause laser radiation or the sun This can have an unwanted effect on the accuracy of the camera It can also cause damage to the detector in the camera CAUTION Do not use the camera in tem...

Page 20: ...he battery be comes hot damage can occur to the safety equipment and this can cause more heat damage or ignition of the battery CAUTION Applicability Cameras with one or more batteries Do not put the battery on a fire or increase the temperature of the battery with heat Damage to the battery and injury to persons can occur CAUTION Applicability Cameras with one or more batteries Do not put the bat...

Page 21: ...tery is 15 C to 50 C 5 F to 122 F unless other information is specified in the user documentation or technical data If you operate the battery out of this temperature range it can decrease the performance or the life cycle of the battery CAUTION Applicability Cameras with one or more batteries When the battery is worn apply insulation to the terminals with adhesive tape or equivalent materials be ...

Page 22: ...or caps This includes the compartments for data storage batteries and connectors CAUTION Applicability Cameras with a viewfinder Make sure that the beams from the intensive energy sources do not go into the viewfinder The beams can cause damage to the camera This includes the devices that emit laser radiation or the sun T559954 r AL 37426 37426 en US 8 ...

Page 23: ...lease contact your FLIR Systems representative for more details 3 5 Training To read about infrared training visit http www infraredtraining com http www irtraining com http www irtraining eu 3 6 Documentation updates Our manuals are updated several times per year and we also issue product critical notifi cations of changes on a regular basis To access the latest manuals translations of manuals an...

Page 24: ...ative versions The authoritative version of this publication is English In the event of divergences due to translation errors the English text has precedence Any late changes are first implemented in English T559954 r AL 37426 37426 en US 10 ...

Page 25: ...ter online If you only want to search the knowledgebase for existing questions and answers you do not need to be a registered user When you want to submit a question make sure that you have the following information to hand The camera model The camera serial number The communication protocol or method between the camera and your device for ex ample HDMI Ethernet USB or FireWire T559954 r AL 37426 ...

Page 26: ...lso download the following when applicable for the product Firmware updates for your infrared camera Program updates for your PC Mac software Freeware and evaluation versions of PC Mac software User documentation for current obsolete and historical products Mechanical drawings in dxf and pdf format Cad data models in stp format Application stories Technical datasheets Product catalogs T559954 r AL...

Page 27: ...y and high sensitivity to accurately measure the smallest temperature differences With real time radiometric recording by the camera it is possible to capture fast events on the camera s SD card for further analysis by the supplied analysis software Flexible and feature rich A wide variety of measuring and analysis functions makes the camera flexible for your every need Two programmable buttons pr...

Page 28: ...just the focus correctly Incorrect focus adjustment af fects how the image modes work It also affects the temperature measurement 7 Push the Save button fully down to save an image 8 Go to http support flir com tools and download FLIR Tools 9 Install FLIR Tools on your computer 10 Start FLIR Tools 11 Connect the camera to the computer using a USB cable 12 Import the images into FLIR Tools 13 Selec...

Page 29: ... FLIR Tools Tools include the following Import images from your camera to your computer Apply filters when searching for images Lay out move and resize measurement tools on any infrared image Group and ungroup files Create panoramas by stitching several smaller images into a larger one Create PDF imagesheets of any images of your choice Add headers footers and logos to imagesheets Create PDF Micro...

Page 30: ...ared images and or digital photos 2 Connect your camera to a PC using a USB connector 3 Import the images from the camera into FLIR Tools Tools 4 Do one of the following Create a PDF imagesheet in FLIR Tools Create a PDF report in FLIR Tools Create a non radiometric Microsoft Word report in FLIR Tools Create a radiometric Microsoft Word report in FLIR Tools 5 Send the report to your client as an a...

Page 31: ...elopment purposes As an example the camera can capture and stream very rap id processes which cannot be perceived by the human eye for later processing and anal ysis in the PC software The streamed video is fully radiometric and uncompressed 8 2 System overview 8 2 1 Figure 8 2 2 Explanation 1 PC running FLIR ResearchIR Max a download card with a printed license key is in cluded with the HSI box 2...

Page 32: ...tart up dialog box click the camera you want to connect to For more information about the installation and connection processes see the FLIR Re searchIR Max manual 8 4 HSI box indicator LED Indicator LED status Explanation The indicator LED displays a continuous blue light The HSI box is operational and video data is now continuously streamed to the computer The indicator LED displays a flashing b...

Page 33: ...amples on how to hold the camera Note Always tilt the touch screen LCD to suit your work position When you hold the camera make sure that you support the optics housing with your left hand too This decreases the strain on your right hand Always use the supplied neck strap See section 10 5 Neck strap attachment points page 25 9 2 Figure T559954 r AL 37426 37426 en US 19 ...

Page 34: ...wfinder 2 Hand strap 3 Digital zoom button 4 Save button push fully down Note The infrared camera can be configured to autofocus when you push the Save button half way down To enable the autofocus function of the Save button select Settings Save options storage Save button half press Autofocus T559954 r AL 37426 37426 en US 20 ...

Page 35: ...nation 1 Digital camera 2 Camera lamp 3 Laser pointer Note The laser pointer may not be enabled in all markets 4 Infrared lens 5 Camera lamp 6 Digital camera 7 Focusing ring 8 Programmable button 9 Button to operate the laser pointer T559954 r AL 37426 37426 en US 21 ...

Page 36: ...ing Do not connect the camera to a computer when this LED is flashing 3 USB Micro B cable to connect the camera to a computer 4 HDMI cable for digital video output 5 Battery condition LED indicator 6 Power supply cable to power the camera and charge the battery 7 High speed interface cable 8 Tripod mount 9 Button to release the lens 10 Latch to release the battery T559954 r AL 37426 37426 en US 22...

Page 37: ...tion Move the joystick left right or up down to navigate in menus submenus and dialog boxes and to change values in dialog boxes Push the joystick to confirm changes and settings in menus and dialog boxes 6 Back button Function Push to leave dialog boxes and to go back into the menu system 7 Camera lamp button Function Push to turn on or off the camera lamp Note The camera lamp must be enabled Sel...

Page 38: ... button Function Push to open the image archive Push and hold for more than 2 seconds to perform a manual calibration 10 Touch screen LCD 11 Power indicator 12 On off button Function Push and release to turn on the camera Push and hold for more than 0 5 second to turn off the camera T559954 r AL 37426 37426 en US 24 ...

Page 39: ...re 10 6 Battery condition LED indicator 10 6 1 Figure 10 6 2 Explanation Type of signal Explanation The green LED flashes twice per second The battery is being charged The green LED glows continuously The battery is fully charged T559954 r AL 37426 37426 en US 25 ...

Page 40: ...ra is off The LED is blue The camera is on 10 8 Laser pointer 10 8 1 Figure Figure 10 1 This figure shows the difference in position between the laser pointer and the optical center of the infrared lens WARNING Do not look directly into the laser beam The laser beam can cause eye irritation T559954 r AL 37426 37426 en US 26 ...

Page 41: ...er Note The laser pointer may not be enabled in all markets 10 8 2 Laser warning label A laser warning label with the following information is attached to the camera 10 8 3 Laser rules and regulations Wavelength 635 nm Maximum output power 1 mW This product complies with 21 CFR 1040 10 and 1040 11 except for deviations pursuant to Laser Notice No 50 dated June 24 2007 T559954 r AL 37426 37426 en U...

Page 42: ...ettings button 11 Color button 12 Measurement button 13 Image mode button 14 Measurement parameters button 15 Recording mode button 11 3 Status icons and indicators Battery status indicator Memory card storage status indicator Manual adjustment mode is enabled The camera lamp is turned on The camera is connected to a device using USB Wi Fi connectivity indicator Bluetooth connectivity indicator T5...

Page 43: ...1 4 Image overlay information The image information consists of items such as date emissivity and atmospheric tem perature All image information is saved in the image file and can be viewed in the image archive You can also choose to display selected items as image overlay information All image overlay information displayed on the live image will also be displayed on saved im ages For more informa...

Page 44: ...s assumed that the joystick is used but most tasks can also be carried out using the index finger or a stylus pen 12 2 Navigating using the joystick You navigate the menu system by using the joystick and the Back button To display the menu system push the joystick To navigate in menus submenus and dialog boxes and to change values in dialog boxes move the joystick up down or left right To confirm ...

Page 45: ...ion The blue LED flashes The battery is being charged The blue LED glows continuously The battery is fully charged 13 1 2 2 Procedure Follow this procedure 1 Put the battery in the battery charger 2 Connect the power supply cable plug to the connector on the battery charger 3 Connect the power supply mains electricity plug to a mains socket 4 It is good practice to disconnect the power supply cabl...

Page 46: ...damage to the camera This includes the devices that emit laser radiation or the sun 13 4 2 Procedure Follow this procedure 1 Push the button to switch from touch screen LCD mode to viewfinder mode 2 To adjust the viewfinder s dioptric correction look through the viewfinder and rotate the adjustment knob clockwise or counter clockwise for the best sharpness Note Maximum dioptric correction 2 Minimu...

Page 47: ...ally 13 6 1 Figure 13 6 2 Procedure Follow this procedure 1 Do one of the following For far focus rotate the focus ring clockwise looking at the touch screen LCD side of the camera For near focus rotate the focus ring counter clockwise looking at the touch screen LCD side of the camera T559954 r AL 37426 37426 en US 33 ...

Page 48: ...aller but faster rotation Furthermore for a very slow rotation the lens moves in very small discrete steps which you can hear allowing for a controlled fine adjust ment of the focus 13 7 Autofocusing the infrared camera 13 7 1 General The infrared camera can be configured to autofocus when you push the Save button half way down 13 7 2 Figure 13 7 3 Procedure Follow this procedure 1 To enable the a...

Page 49: ...h the joystick to display the menu system 2 Use the joystick to go to Settings 3 Push the joystick to display the Settings menu 4 Select Device settings and push the joystick 5 Select Continuous autofocus and push the joystick 6 Select On and push the joystick Note You can also assign the function Continuous autofocus to one of the program mable buttons Select Settings Programmable buttons 13 9 Op...

Page 50: ... laser pointer release the laser button Note The symbol is displayed on the screen when the laser pointer is on 13 10 Using the digital zoom function 13 10 1 General The current zoom factor is displayed in the upper left corner of the screen In preview edit mode it is possible to pan a zoomed image by touching the screen T559954 r AL 37426 37426 en US 36 ...

Page 51: ...ow this procedure 1 To zoom push the zoom button left or right 13 11 Assigning functions to the programmable buttons 13 11 1 General There are four programmable buttons two hardware buttons and two software buttons on the screen T559954 r AL 37426 37426 en US 37 ...

Page 52: ...sh the joystick 13 12 Using the camera lamp as a flash 13 12 1 General The camera lamp can be used as a flash for the digital camera When the flash function is activated the camera lamp will flash when an image is saved by pushing the Save button fully down 13 12 2 Procedure Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Settings 3 Push the joystic...

Page 53: ...Handling the camera 13 Follow this procedure 1 Take a firm grip around the outermost part of the lens 2 Push the release button and rotate the lens 45 counter clockwise T559954 r AL 37426 37426 en US 39 ...

Page 54: ...ll out the lens from the lens bayonet mount 4 The infrared detector is now fully exposed Do not touch this surface If you see dust on the detector follow the instructions in 31 3 Infrared detector page 217 T559954 r AL 37426 37426 en US 40 ...

Page 55: ...dling the camera 13 5 Note the index marks on the lens bayonet mount and on the lens 6 Align the lens correctly to the bayonet mount Carefully push the lens into position T559954 r AL 37426 37426 en US 41 ...

Page 56: ...ns see section 13 14 2 Attaching the close up lens page 42 For best performance adjust the focus to infinity The working distance for the close up lens is 97 mm The working distance is the dis tance between the front of the lens and the closest surface of the object when the ob ject is in sharp focus You must activate the global parameter External IR window compensation Only use the Thermal image ...

Page 57: ...r adjust the focus manually see section 13 6 Adjusting the infrared camera focus manually page 33 Note Once the focus is set to infinity be careful when you continue with the proce dure Make sure you do not rotate the focusing ring by mistake 2 Remove the outermost rubber protection from the infrared lens 3 Align the close up lens with the infrared lens Carefully push the close up lens into positi...

Page 58: ... Push the Back button to go back in the menu system 5 6 The status icon is now displayed 6 Select the image mode Thermal by doing the following 6 1 Push the joystick to display the menu system 6 2 Select Image mode and push the joystick This displays a submenu 6 3 Select Thermal and push the joystick 6 4 Push the Back button to go back in the menu system 7 Deactivate continuous autofocus by select...

Page 59: ...ull out the close up lens from the infrared lens 3 Note the grooves on the infrared lens and their counterparts on the inside of the rubber protection 4 Align the rubber protection with the infrared lens T559954 r AL 37426 37426 en US 45 ...

Page 60: ...compass It is recommended that the compass is calibrated every time you move the camera to a new location 13 15 1 Procedure Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Settings 3 Push the joystick to display the Settings menu 4 Select Device settings and push the joystick 5 Depending on the camera configuration select Wireless geolocation Wirele...

Page 61: ...tions storage page 83 14 1 3 Image capacity The capacity of a 4 GB memory card is theoretically 1000 images with no annotations 14 1 4 About UltraMax UltraMax is an image enhancement feature that increases the image resolution and lowers the noise making small objects easier to see and measure An UltraMax image is twice as wide and high as an ordinary image When an UltraMax image is captured by th...

Page 62: ...s procedure 1 To preview an image push the Save button fully down This displays the preview 2 Manual image adjustment mode is now active and the status icon is displayed For image adjustment instructions see section 15 3 Adjusting the infrared image page 53 3 To edit the image push the joystick This displays a context menu For editing instruc tions see section 14 5 Editing a saved image page 49 4 ...

Page 63: ...age adjustment instructions see section 15 3 Adjusting the infrared image page 53 4 Push the joystick This displays a context menu Select Cancel to exit edit mode Select Measurement parameters to change the global parameters Select Image mode to change the image mode Select Measurement to add a measurement tool Select Color to change the color palette or set a color alarm Select Save to save and e...

Page 64: ... an image 14 7 1 General You can delete an image file from the memory card Note Both images in the image file thermal and visual will be deleted 14 7 2 Procedure Follow this procedure 1 Push the Image archive button 2 Move the joystick up down or left right to select the image you want to delete 3 Push the joystick to display the image 4 Push the joystick and select Delete from the menu This displ...

Page 65: ...k to go to Settings 3 Push the joystick to display the Settings menu 4 Select Save options storage and push the joystick 5 Select Delete all saved files and push the joystick This displays a dialog box where you can choose to execute the delete action or to cancel 6 To permanently delete all saved files select Delete and push the joystick 14 10 Resetting the image counter 14 10 1 General You can r...

Page 66: ... Select Device settings and push the joystick 5 Select Reset options and push the joystick 6 Select Reset image counter and push the joystick This displays a dialog box where you can choose to execute the reset action or to cancel 7 To reset the counter select Reset and push the joystick T559954 r AL 37426 37426 en US 52 ...

Page 67: ...us function to the programmable button or For more information see section 13 11 Assigning functions to the programmable buttons page 37 The infrared camera can also be set up to perform continuous autofocusing For more in formation see section 13 8 Continuous autofocus page 35 15 3 Adjusting the infrared image 15 3 1 General An infrared image can be adjusted automatically or manually In automatic...

Page 68: ... to perform an automatic adjustment of the image while remaining in manual image adjustment mode Select Settings Pro grammable buttons 15 3 2 Example 1 Here are two infrared images of a building In the left image which is auto adjusted the large temperature span between the clear sky and the heated building makes a correct analysis difficult You can analyze the building in more detail if you chang...

Page 69: ...face options Manual adjustment mode Level Max Min Follow this procedure 1 In live mode push the button to enter manual adjustment mode 2 To simultaneously change the temperature scale minimum and maximum limits move the joystick up down 3 To change the minimum limit or the maximum limit do the following Move the joystick left right to select highlight the maximum or minimum temperature Move the jo...

Page 70: ... range to one of the program mable buttons Select Settings Programmable buttons 15 5 Changing the color palette 15 5 1 General You can change the color palette that the camera uses to display different temperatures A different palette can make it easier to analyze an image Iron Arctic Rainbow Rainbow high contrast T559954 r AL 37426 37426 en US 56 ...

Page 71: ...6 Changing the object parameters For accurate measurements you must set the object parameters External IR window compensation Object distance Atmospheric temperature Relative humidity Reflected temperature Emissivity You can set the object parameters globally You can also change the Emissivity Reflected temperature and Object distance parameters locally for a measurement tool For more information ...

Page 72: ...cording a video sequence 15 7 2 1 Procedure Follow this procedure 1 To perform a manual calibration push and hold down the Image archive button for more than 2 seconds Note You can also assign the function Calibrate to one of the programmable buttons Se lect Settings Programmable buttons 15 8 Hiding all overlay 15 8 1 General The camera overlay consists of overlay graphics and image overlay inform...

Page 73: ...to go to Settings 3 Push the joystick to display the Settings menu 4 Select Programmable buttons and push the joystick 5 Select one of the buttons and push the joystick 6 Select Hide image overlay graphics and push the joystick You have now assigned this function to the selected programmable button T559954 r AL 37426 37426 en US 59 ...

Page 74: ...on top of the visual image Digital camera The visual image captured by the digital camera is displayed Note These image modes only work correctly for calibrated lenses The lens that ships with the camera is factory calibrated To have a new lens calibrated you must send in the camera and the lens to your local service department All thermal and visual information is stored when an image is saved Th...

Page 75: ...oystick to go to one of the image modes Thermal MSX Thermal Picture in picture Digital camera Note If the csq video format is selected Settings Save options storage Video compression and the recording mode Video is selected it will only be possible to se lect the image mode Thermal 5 Push the joystick confirm 6 If Picture in picture mode is selected you can at this point move and resize the infrar...

Page 76: ... to add a hot spot detection within a box area Select Cold spot to add a cold spot detection within a box area Select User preset 1 to add user preset 1 Not available in all camera models Select User preset 2 to add user preset 2 Not available in all camera models 4 Push the joystick This displays the measurement tool or the group of preset tools on the screen 17 3 Working with user presets 17 3 1...

Page 77: ...isplays the measurement tool on the screen 9 Push the joystick This displays a context menu where you can select one or more of the following actions depending on the type of tool Remove the tool Resize move center and rotate the tool Display maximum minimum and average values Set alarms Set local parameters When completed select Done 10 When all measurement tools have been added select Save as pr...

Page 78: ...ject of interest Atmospheric temperature i e the temperature of the air between the camera and the object of interest Relative humidity i e the relative humidity of the air between the camera and the ob ject of interest Reflected temperature which is used when compensating for the radiation from the surroundings reflected by the object into the camera This property of the object is called reflecti...

Page 79: ... global pa rameters are usually sufficient Note Of the object parameters Emissivity and Reflected temperature are the two most important to set correctly in the camera 17 5 4 1 Setting global parameters Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Measurement parameters 3 Push the joystick to display a submenu Use the joystick to select one or mo...

Page 80: ...u 6 Use the joystick to select an object parameter 7 Push the joystick to display a dialog box 8 Use the joystick to change the parameter 9 Push the joystick This closes the dialog box 17 5 5 Related topics For in depth information about parameters and how to correctly set the emissivity and re flected apparent temperature see section 35 Thermographic measurement techniques page 229 17 6 Displayin...

Page 81: ...ystick to display the average value Optional step You can choose to show or hide the maximum and minimum markers the hot cold spots Select Max min markers and push the joystick to toggle When icon with grey indicator is displayed the markers are hidden When icon with blue indicator is displayed the markers are shown 5 When completed move the joystick down to close the submenu 6 Select Done and pus...

Page 82: ...sult of the difference calculation is now displayed on the screen 17 8 Setting a measurement alarm 17 8 1 General You can make the camera trigger an alarm when certain measurement conditions are met 17 8 2 Types of alarm You can choose between the following alarm types Above Triggers an alarm when the temperature is above the preset alarm temperature Below Triggers an alarm when the temperature is...

Page 83: ...a to display at least one value maximum minimum or average in the result table Follow this procedure 1 To select the measurement tool touch the tool on the screen The tool is now dis played with one or more handles 2 Push the joystick or touch and hold the tool This displays a context menu 3 Use the joystick to go to depending on tool or Max Min Avg Alarm or Graph Max Min Avg Alarm 4 Push the joys...

Page 84: ...alog box 3 Select Delta This displays a context menu 4 Use the joystick to go to Set alarm on delta 5 Push the joystick This displays a dialog box where you can define the settings for the alarm Alarm condition The condition that triggers the alarm Applicable values are Above Below or Off Alarm limit The temperature value that will be the critical condition when an alarm is triggered or not Alarm ...

Page 85: ...his will apply a contrasting color to all pixels with a temperature above the specified temperature level Below alarm This will apply a contrasting color to all pixels with a temperature below the specified temperature level Interval alarm This will apply a contrasting color to all pixels with a temperature be tween two specified temperature levels Condensation alarm Triggers when the camera detec...

Page 86: ... alarms Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Color 3 Push the joystick to display a submenu Use the joystick to select the type of alarm Above alarm Below alarm Interval alarm 4 Push the joystick The threshold temperature is displayed at the top of the screen T559954 r AL 37426 37426 en US 72 ...

Page 87: ...uildings Refer to your national building code for recommendations 18 3 3 Setting up condensation and insulation alarms Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Color 3 Push the joystick to display a submenu Use the joystick to select the type of alarm Condensation alarm Insulation alarm 4 Push the joystick This displays a dialog box where you...

Page 88: ...dding annotations to a saved image in the image archive Adding annotations when saving an image works in a similar way 19 2 Adding a note 19 2 1 General You can add a text note to the image file Using this feature you can annotate images by entering free form text 19 2 2 Procedure Follow this procedure 1 Open the image in the image archive 2 Push the joystick This displays a context menu 3 Select ...

Page 89: ...plate by first selecting Select default template 5 For each row in the table do the following Push the joystick This displays the predefined values Move the joystick up down to select a predefined value Push the joystick to confirm Instead of selecting a predefined value you can select the keyboard and enter other text by touching the screen 6 When completed select Save Exit at the bottom of the t...

Page 90: ...o listen to the recording select Play and push the joystick 7 To delete the recording select Delete and push the joystick 8 When completed select Done and push the joystick 19 5 Adding a sketch 19 5 1 General You can add a freehand drawing to an image 19 5 2 Procedure Follow this procedure 1 Open the image in the image archive 2 Push the joystick This displays a context menu 3 Select Add sketch an...

Page 91: ...the joystick Erase parts of the sketch by touching the screen To add an arrow circle or cross select Stamp sketch and push the joystick Select the type of stamp and push the joystick The stamp is displayed in the center of the screen You can move the stamp by using the joystick or by touching the screen To clear select Clear all and push the joystick When the sketch is completed select Save and pu...

Page 92: ...t Time lapse 5 Push the joystick This displays a dialog box where you can set the save conditions Save interval Use the joystick to set the time interval between each saved image Total number of images Periodic saving will stop when the set number of images have been saved 6 Push the joystick This closes the dialog box The time interval is displayed at the top of the screen 7 To manually start or ...

Page 93: ...e is supported when recording video If any other image mode is active when Video recording mode is selected the camera will auto switch to Thermal image mode 21 2 Procedure Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Recording mode 3 Push the joystick This displays a submenu 4 Select Video and push the joystick 5 Do the following To start a reco...

Page 94: ...ecording mode 4 Push the joystick This displays a submenu 5 Select Screening 6 Push the joystick This displays a dialog box where you can define the settings for the alarm Allowed deviation The allowed deviation from the sampled average Alarm sound Applicable values are Beep or No sound 7 Push the joystick This closes the dialog box 8 Aim the camera toward a point of interest The object should be ...

Page 95: ...box is unchecked push the joystick to activate Bluetooth Note You also need to ensure that the external Bluetooth device is in visible mode 8 Select Available devices and push the joystick 9 Wait until a list of available devices is displayed This will take about 15 seconds 10 When a Bluetooth device is found select the device to add it and begin the pairing procedure The device is then ready to b...

Page 96: ...joystick To change the WPA2 password select Password and push the joystick Note These parameters are set for your camera s network They will be used by the external device to connect that device to the network 24 3 Connecting the camera to a WLAN less common use Follow this procedure 1 Push the joystick to display the menu system 2 Use the joystick to go to Settings 3 Push the joystick to display ...

Page 97: ... are inspecting Available temperature range options are dependent on the camera model The unit or depends on the temperature unit setting see section 25 1 5 Device settings page 85 25 1 3 Save options storage Save button half press This setting defines the function of the Save button Available options are Autofocus Pushing the Save button half way down will autofocus the infrared camera None Pushi...

Page 98: ... setting has no impact on already saved files in the archive Available options are DCF DCF Design rule for Camera File system is a standard that specifies the nam ing method of image files and much more With this setting the name of a saved image video file will be FLIRxxxx where xxxx is a unique counter Example FLIR0001 Date prefix A prefix will be added to the file name including the date and th...

Page 99: ...g Lamp laser is set to the option Disable all For more information see section 25 1 5 Device settings page 85 Switch single shot Video Switch between the recording modes Single shot and Video Switch between two latest palettes Switch between the two last used color palettes For more information see section 15 5 Changing the color palette page 56 Switch temperature range Cycle through the camera te...

Page 100: ... Wi Fi page 82 Bluetooth This setting defines Bluetooth connectivity For more information see section 23 Pairing Bluetooth devices page 81 GPS This setting is used to enable disable the GPS Compass This setting is used to enable disable the compass and to calibrate the compass For more information see section 13 15 Calibrating the compass page 46 Lamp laser This submenu includes the following sett...

Page 101: ... Note When a reset option is selected a dialog box is displayed with more information You can choose to execute the reset action or to cancel Camera information This dialog displays information about the camera No changes can be made Model Serial number Part number Software version Storage The used and free space on the memory card Lens The field of view of the lens Battery The remaining battery c...

Page 102: ... data FLIR Systems reserves the right to change specifications at any time without prior notice Please check http support flir com for latest changes 26 3 Note about authoritative versions The authoritative version of this publication is English In the event of divergences due to translation errors the English text has precedence Any late changes are first implemented in English T559954 r AL 37426...

Page 103: ...n enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 12 9 Minimum IR focus distance 1 3 m 4 26 ft Minimum IR visual alignment distance 1 3 m 4 26 ft Focal length 8...

Page 104: ...02 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marke...

Page 105: ...pass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas u...

Page 106: ...video streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automat...

Page 107: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 108: ...y charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B EAN 13 7332558010273 UPC 12 845188010911 Country of origin Sweden Supplies accessories T199064 IR lens f 36mm 28 with case T199066 IR lens f 21 2mm 45 with case T199077 IR ...

Page 109: ...IR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T559954 r AL 37426 37426 en US 95 ...

Page 110: ...n enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 28 21 Minimum IR focus distance 0 4 m 1 32 ft Minimum IR visual alignment distance 0 4 m 1 32 ft Focal length ...

Page 111: ...02 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marke...

Page 112: ...pass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas u...

Page 113: ...video streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automat...

Page 114: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 115: ...y charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B EAN 13 7332558010280 UPC 12 845188010928 Country of origin Sweden Supplies accessories T199064 IR lens f 36mm 28 with case T199066 IR lens f 21 2mm 45 with case T199077 IR ...

Page 116: ...R Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T559954 r AL 37426 37426 en US 102 ...

Page 117: ... enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 45 34 Minimum IR focus distance 0 2 m 0 66 ft Minimum IR visual alignment distance 0 5 m 1 64 ft Focal length 2...

Page 118: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 119: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 120: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 121: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 122: ...y charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B EAN 13 7332558010297 UPC 12 845188010935 Country of origin Sweden Supplies accessories T199064 IR lens f 36mm 28 with case T199066 IR lens f 21 2mm 45 with case T199077 IR ...

Page 123: ...R Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T559954 r AL 37426 37426 en US 109 ...

Page 124: ...When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 28 21 Minimum IR focus distance 0 4 m 1 32 ft Minimum IR visual alignment distance 0 4 m 1 32 ft Focal leng...

Page 125: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 126: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 127: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 128: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 129: ...rd box List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable IR lens 12 Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B Packaging weight 7 2 kg 15 8 lb Packaging size 494 355 188 mm 19 4 14 0 7 4 in EAN 13 733255...

Page 130: ...ool belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T55995...

Page 131: ...When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 28 21 Minimum IR focus distance 0 4 m 1 32 ft Minimum IR visual alignment distance 0 4 m 1 32 ft Focal leng...

Page 132: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 133: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 134: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 135: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 136: ...rd box List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable IR lens 45 Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B Packaging weight 6 9 kg 15 2 lb Packaging size 494 355 188 mm 19 4 14 0 7 4 in EAN 13 733255...

Page 137: ...ool belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T55995...

Page 138: ...x When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 28 21 Minimum IR focus distance 0 4 m 1 32 ft Minimum IR visual alignment distance 0 4 m 1 32 ft Focal le...

Page 139: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 140: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 141: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 142: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 143: ...ype Cardboard box List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable IR lens 12 IR lens 45 Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B Packaging weight 8 2 kg 18 0 lb Packaging size 494 355 188 mm 19 4 14 ...

Page 144: ...Headset T911093 Tool belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK ...

Page 145: ...t performance with the latest technology The FLIR T1030sc is equipped with the innovative Multi Spectral Dynamic Imaging MSX feature which produces an image richer in detail than ever be fore With its continuous autofocus the FLIR T1030sc is a fully automatic infrared camera Support for UltraMax When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when impor...

Page 146: ...range 40 to 150 C 40 to 302 F 0 to 650 C 32 to 1202 F 300 to 2000 C 572 to 3632 F Accuracy 1 C 1 8 F or 1 25 C 77 F for tem peratures between 5 and 150 C 41 and 302 F in range 40 to 150 C 40 to 302 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI high speed interface use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of...

Page 147: ...n alarm Audible visual alarms above below on any se lected measurement function Set up Set up commands Define user presets Save options Programmable button Reset options Set up camera Wi Fi GPS compass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simpli...

Page 148: ...aming Full dynamic un compressed 120 Hz 16 bit full frame 2 Gbit to a PC using an HSI box Real time radiometric streaming 30 Hz RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Windowing 30 Hz 1024 768 full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for ...

Page 149: ...quency range 2402 2480 MHz supports 2 1 4 0 and 4 0 BLE Antenna Internal Power system Battery type Rechargeable Li ion battery Battery operating time 2 5 hours at 25 C 68 F and typical use Charging system In camera AC adapter or 12 V from a vehicle or two bay charger Charging time 2 5 hours to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operat...

Page 150: ... 6 6 in 8 0 in 7 4 in Tripod mounting UNC 20 Housing material Magnesium Warranty information Warranty 2 years parts and labor coverage on the camera 5 years coverage on the battery 10 years coverage on the detector the most vital part of the whole camera Shipping information List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools...

Page 151: ...l belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T198697 FLIR ResearchIR Max HSDR 4 hardware sec dev T199014 FLIR Research...

Page 152: ...t performance with the latest technology The FLIR T1030sc is equipped with the innovative Multi Spectral Dynamic Imaging MSX feature which produces an image richer in detail than ever be fore With its continuous autofocus the FLIR T1030sc is a fully automatic infrared camera Support for UltraMax When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when impor...

Page 153: ...range 40 to 150 C 40 to 302 F 0 to 650 C 32 to 1202 F 300 to 2000 C 572 to 3632 F Accuracy 1 C 1 8 F or 1 25 C 77 F for tem peratures between 5 and 150 C 41 and 302 F in range 40 to 150 C 40 to 302 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI high speed interface use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of...

Page 154: ...n alarm Audible visual alarms above below on any se lected measurement function Set up Set up commands Define user presets Save options Programmable button Reset options Set up camera Wi Fi GPS compass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simpli...

Page 155: ...aming Full dynamic un compressed 120 Hz 16 bit full frame 2 Gbit to a PC using an HSI box Real time radiometric streaming 30 Hz RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Windowing 30 Hz 1024 768 full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for ...

Page 156: ...quency range 2402 2480 MHz supports 2 1 4 0 and 4 0 BLE Antenna Internal Power system Battery type Rechargeable Li ion battery Battery operating time 2 5 hours at 25 C 68 F and typical use Charging system In camera AC adapter or 12 V from a vehicle or two bay charger Charging time 2 5 hours to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operat...

Page 157: ... 6 6 in 8 0 in 7 4 in Tripod mounting UNC 20 Housing material Magnesium Warranty information Warranty 2 years parts and labor coverage on the camera 5 years coverage on the battery 10 years coverage on the detector the most vital part of the whole camera Shipping information List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools...

Page 158: ...l belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T198697 FLIR ResearchIR Max HSDR 4 hardware sec dev T199014 FLIR Research...

Page 159: ...t performance with the latest technology The FLIR T1030sc is equipped with the innovative Multi Spectral Dynamic Imaging MSX feature which produces an image richer in detail than ever be fore With its continuous autofocus the FLIR T1030sc is a fully automatic infrared camera Support for UltraMax When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when impor...

Page 160: ...range 40 to 150 C 40 to 302 F 0 to 650 C 32 to 1202 F 300 to 2000 C 572 to 3632 F Accuracy 1 C 1 8 F or 1 25 C 77 F for tem peratures between 5 and 150 C 41 and 302 F in range 40 to 150 C 40 to 302 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI high speed interface use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of...

Page 161: ...n alarm Audible visual alarms above below on any se lected measurement function Set up Set up commands Define user presets Save options Programmable button Reset options Set up camera Wi Fi GPS compass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simpli...

Page 162: ...aming Full dynamic un compressed 120 Hz 16 bit full frame 2 Gbit to a PC using an HSI box Real time radiometric streaming 30 Hz RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Windowing 30 Hz 1024 768 full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for ...

Page 163: ...quency range 2402 2480 MHz supports 2 1 4 0 and 4 0 BLE Antenna Internal Power system Battery type Rechargeable Li ion battery Battery operating time 2 5 hours at 25 C 68 F and typical use Charging system In camera AC adapter or 12 V from a vehicle or two bay charger Charging time 2 5 hours to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operat...

Page 164: ... 6 6 in 8 0 in 7 4 in Tripod mounting UNC 20 Housing material Magnesium Warranty information Warranty 2 years parts and labor coverage on the camera 5 years coverage on the battery 10 years coverage on the detector the most vital part of the whole camera Shipping information List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools...

Page 165: ...l belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T198697 FLIR ResearchIR Max HSDR 4 hardware sec dev T199014 FLIR Research...

Page 166: ...n enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 12 9 Minimum IR focus distance 1 3 m 4 26 ft Minimum IR visual alignment distance 1 3 m 4 26 ft Focal length 8...

Page 167: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 168: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 169: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 170: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 171: ...y charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B EAN 13 7332558010389 UPC 12 845188011031 Country of origin Sweden Supplies accessories T199064 IR lens f 36mm 28 with case T199066 IR lens f 21 2mm 45 with case T199077 IR ...

Page 172: ...R Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T559954 r AL 37426 37426 en US 158 ...

Page 173: ...n enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 28 21 Minimum IR focus distance 0 4 m 1 32 ft Minimum IR visual alignment distance 0 4 m 1 32 ft Focal length ...

Page 174: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 175: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 176: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 177: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 178: ...y charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B EAN 13 7332558010396 UPC 12 845188011048 Country of origin Sweden Supplies accessories T199064 IR lens f 36mm 28 with case T199066 IR lens f 21 2mm 45 with case T199077 IR ...

Page 179: ...R Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T559954 r AL 37426 37426 en US 165 ...

Page 180: ...n enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when importing the images into FLIR Tools Imaging and optical data IR resolution 1024 768 pixels MSX resolution 1024 768 pixels UltraMax Yes Thermal sensitivity NETD 20 mK 30 C 86 F Field of view FOV 45 34 Minimum IR focus distance 0 2 m 0 66 ft Minimum IR visual alignment distance 0 5 m 1 64 ft Focal length ...

Page 181: ...2 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of reading 25 C 77 F Measurement analysis Spotmeter 10 Area 5 5 areas boxes and circles with max min average Profile 1 line profile with max min temperature Automatic hot cold detection Auto hot or cold spotmeter marker...

Page 182: ...ass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simplified Chinese Swedish traditional Chinese Turkish Service functions Camera software update Use PC software FLIR Tools Storage of images Image storage Standard JPEG including digital image and meas ur...

Page 183: ...ideo streaming Real time radiometric streaming RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Digital camera Digital camera Field of view adapts to the infrared lens Video lamp Built in LED light Laser pointer Laser Activated by a dedicated button Laser alignment Position is automati...

Page 184: ...rs to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operation AC adapter 90 260 V AC 50 60 Hz or 12 V from a vehicle cable with a standard plug optional Power management Automatic power off functionality user configura ble between 5 minutes 20 minutes and no auto matic shutdown Environmental data Operating temperature range 15 C to 50 C 5 F to 1...

Page 185: ...y charger Bluetooth headset Calibration certificate FLIR Tools license card Hard transport case HDMI HDMI cable Lens cap Memory card Neck strap Power supply including multi plugs Printed documentation USB cable Std A to Micro B EAN 13 7332558010402 UPC 12 845188011055 Country of origin Sweden Supplies accessories T199064 IR lens f 36mm 28 with case T199066 IR lens f 21 2mm 45 with case T199077 IR ...

Page 186: ...R Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T199233 FLIR Atlas SDK for NET T199234 FLIR Atlas SDK for MATLAB T559954 r AL 37426 37426 en US 172 ...

Page 187: ...t performance with the latest technology The FLIR T1050sc is equipped with the innovative Multi Spectral Dynamic Imaging MSX feature which produces an image richer in detail than ever be fore With its continuous autofocus the FLIR T1050sc is a fully automatic infrared camera Support for UltraMax When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when impor...

Page 188: ...range 40 to 150 C 40 to 302 F 0 to 650 C 32 to 1202 F 300 to 2000 C 572 to 3632 F Accuracy 1 C 1 8 F or 1 25 C 77 F for tem peratures between 5 and 150 C 41 and 302 F in range 40 to 150 C 40 to 302 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI high speed interface use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of...

Page 189: ...n alarm Audible visual alarms above below on any se lected measurement function Set up Set up commands Define user presets Save options Programmable button Reset options Set up camera Wi Fi GPS compass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simpli...

Page 190: ...aming Full dynamic un compressed 120 Hz 16 bit full frame 2 Gbit to a PC using an HSI box Real time radiometric streaming 30 Hz RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Windowing 30 Hz 1024 768 full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for ...

Page 191: ...quency range 2402 2480 MHz supports 2 1 4 0 and 4 0 BLE Antenna Internal Power system Battery type Rechargeable Li ion battery Battery operating time 2 5 hours at 25 C 68 F and typical use Charging system In camera AC adapter or 12 V from a vehicle or two bay charger Charging time 2 5 hours to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operat...

Page 192: ... 6 6 in 8 0 in 7 4 in Tripod mounting UNC 20 Housing material Magnesium Warranty information Warranty 2 years parts and labor coverage on the camera 5 years coverage on the battery 10 years coverage on the detector the most vital part of the whole camera Shipping information List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools...

Page 193: ...l belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T198697 FLIR ResearchIR Max HSDR 4 hardware sec dev T199014 FLIR Research...

Page 194: ...t performance with the latest technology The FLIR T1050sc is equipped with the innovative Multi Spectral Dynamic Imaging MSX feature which produces an image richer in detail than ever be fore With its continuous autofocus the FLIR T1050sc is a fully automatic infrared camera Support for UltraMax When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when impor...

Page 195: ...range 40 to 150 C 40 to 302 F 0 to 650 C 32 to 1202 F 300 to 2000 C 572 to 3632 F Accuracy 1 C 1 8 F or 1 25 C 77 F for tem peratures between 5 and 150 C 41 and 302 F in range 40 to 150 C 40 to 302 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI high speed interface use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of...

Page 196: ...n alarm Audible visual alarms above below on any se lected measurement function Set up Set up commands Define user presets Save options Programmable button Reset options Set up camera Wi Fi GPS compass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simpli...

Page 197: ...aming Full dynamic un compressed 120 Hz 16 bit full frame 2 Gbit to a PC using an HSI box Real time radiometric streaming 30 Hz RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Windowing 30 Hz 1024 768 full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for ...

Page 198: ...quency range 2402 2480 MHz supports 2 1 4 0 and 4 0 BLE Antenna Internal Power system Battery type Rechargeable Li ion battery Battery operating time 2 5 hours at 25 C 68 F and typical use Charging system In camera AC adapter or 12 V from a vehicle or two bay charger Charging time 2 5 hours to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operat...

Page 199: ... 6 6 in 8 0 in 7 4 in Tripod mounting UNC 20 Housing material Magnesium Warranty information Warranty 2 years parts and labor coverage on the camera 5 years coverage on the battery 10 years coverage on the detector the most vital part of the whole camera Shipping information List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools...

Page 200: ...l belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T198697 FLIR ResearchIR Max HSDR 4 hardware sec dev T199014 FLIR Research...

Page 201: ...t performance with the latest technology The FLIR T1050sc is equipped with the innovative Multi Spectral Dynamic Imaging MSX feature which produces an image richer in detail than ever be fore With its continuous autofocus the FLIR T1050sc is a fully automatic infrared camera Support for UltraMax When enabling UltraMax in the camera the resolution of images can be sub stantially enhanced when impor...

Page 202: ...range 40 to 150 C 40 to 302 F 0 to 650 C 32 to 1202 F 300 to 2000 C 572 to 3632 F Accuracy 1 C 1 8 F or 1 25 C 77 F for tem peratures between 5 and 150 C 41 and 302 F in range 40 to 150 C 40 to 302 F 2 C 3 6 F or 2 of reading 25 C 77 F for temperatures between 40 and 2000 C 40 and 3632 F NOTE For HSI high speed interface use above 30 Hz frame rate the typical accuracy will be 2 5 C 4 5 F or 2 5 of...

Page 203: ...n alarm Audible visual alarms above below on any se lected measurement function Set up Set up commands Define user presets Save options Programmable button Reset options Set up camera Wi Fi GPS compass Bluetooth Language Time units Camera information Languages Arabic Czech Danish Dutch English Finnish French German Greek Hungarian Italian Japa nese Korean Norwegian Polish Portuguese Russian simpli...

Page 204: ...aming Full dynamic un compressed 120 Hz 16 bit full frame 2 Gbit to a PC using an HSI box Real time radiometric streaming 30 Hz RTRS via USB Non radiometric IR video streaming H 264 video using Wi Fi H 264 video using USB Visual video streaming H 264 video using Wi Fi H 264 video using USB Windowing 30 Hz 1024 768 full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for ...

Page 205: ...quency range 2402 2480 MHz supports 2 1 4 0 and 4 0 BLE Antenna Internal Power system Battery type Rechargeable Li ion battery Battery operating time 2 5 hours at 25 C 68 F and typical use Charging system In camera AC adapter or 12 V from a vehicle or two bay charger Charging time 2 5 hours to 90 capacity charging status indi cated by LEDs Charging temperature 0 45 C 32 113 F External power operat...

Page 206: ... 6 6 in 8 0 in 7 4 in Tripod mounting UNC 20 Housing material Magnesium Warranty information Warranty 2 years parts and labor coverage on the camera 5 years coverage on the battery 10 years coverage on the detector the most vital part of the whole camera Shipping information List of contents Infrared camera with lens Battery 2 ea Battery charger Bluetooth headset Calibration certificate FLIR Tools...

Page 207: ...l belt T198533 USB cable Std A Micro B T198586 FLIR Reporter Professional license only T198584 FLIR Tools T198583 FLIR Tools download card incl license key DSW 10000 FLIR IR Camera Player APP 10002 FLIR Tools Mobile Android Application APP 10003 FLIR Tools Mobile iPad iPhone Application APP 10004 FLIR Tools MacOS Application T198697 FLIR ResearchIR Max HSDR 4 hardware sec dev T199014 FLIR Research...

Page 208: ...in Shipping information Packaging type Cardboard box List of contents Lens Lens case Front lens cap Rear lens cap Mounting instruction Packaging weight 1 1 kg 2 4 lb Packaging size 173 135 135 mm 6 81 5 31 5 31 in EAN 13 7332558010983 UPC 12 845188011833 Country of origin Sweden Compatible with the following products 72501 0101 FLIR T1020 12 72501 0102 FLIR T1020 28 72501 0103 FLIR T1020 45 72501 ...

Page 209: ...Technical data 26 72501 0402 FLIR T1050sc 28 72501 0403 FLIR T1050sc 45 T559954 r AL 37426 37426 en US 195 ...

Page 210: ...ion IFOV 51 µm Lens identification No F number 1 15 Number of lenses 2 2 asph Distortion 0 4 Physical data Weight 0 620 kg 1 37 lb Size L D 127 5 91 mm 5 02 3 58 in Front lens diameter 76 mm 3 0 in Shipping information Packaging type Cardboard box List of contents Lens Lens case Front lens cap Rear lens cap Mounting instruction Packaging weight 0 96 kg 2 1 lb Packaging size 173 135 135 mm 6 81 5 3...

Page 211: ...0 754 kg 1 66 lb Size L D 96 9 91 mm 3 82 3 58 in Front lens diameter 59 mm 2 3 in Shipping information Packaging type Cardboard box List of contents Lens Lens case Front lens cap Rear lens cap Mounting instruction Packaging weight 1 1 kg 2 4 lb Packaging size 173 135 135 mm 6 81 5 31 5 31 in EAN 13 7332558010990 UPC 12 845188011840 Country of origin Sweden Compatible with the following products 7...

Page 212: ...Technical data 26 72501 0401 FLIR T1050sc 12 72501 0402 FLIR T1050sc 28 72501 0403 FLIR T1050sc 45 T559954 r AL 37426 37426 en US 198 ...

Page 213: ...Size L D 134 1 100 5 mm 5 28 3 96 in Front lens diameter 75 mm 3 0 in Shipping information Packaging type Cardboard box List of contents Lens Lens case Front lens cap Rear lens cap Mounting instruction Packaging weight 1 4 kg 3 1 lb Packaging size 212 150 150 mm 8 35 5 91 5 91 in EAN 13 7332558011010 UPC 12 845188011864 Country of origin Sweden Compatible with the following products 72501 0101 FLI...

Page 214: ...Technical data 26 72501 0401 FLIR T1050sc 12 72501 0402 FLIR T1050sc 28 72501 0403 FLIR T1050sc 45 T559954 r AL 37426 37426 en US 200 ...

Page 215: ... full image height Based on 30 Hz 120 Hz windowing 1024 192 of full image height for range 40 to 150 C 40 to 302 F 120 Hz 1024 768 full image height Based on 120 Hz 240 Hz windowing 1024 384 of full image height for range 0 to 650 C 32 to 1202 F and range 300 to 2000 C 572 to 3632 F Power system Power HSI box Powered from PC via USB 3 Physical data Weight HSI box 0 405 kg 0 893 lb Weight USB cable...

Page 216: ...UPC 12 845188011888 Country of origin Sweden 72501 0201 FLIR T1030sc 12 72501 0202 FLIR T1030sc 28 72501 0203 FLIR T1030sc 45 72501 0401 FLIR T1050sc 12 72501 0402 FLIR T1050sc 28 72501 0403 FLIR T1050sc 45 T559954 r AL 37426 37426 en US 202 ...

Page 217: ...Mechanical drawings 27 T559954 r AL 37426 37426 en US 203 ...

Page 218: ... 7 8 9 10 A B C D E F G H 1 3 2 5 4 C F B D G E A 6 Size D 1 2 Scale 2012 FLIR Systems Inc All rights reserved worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Di...

Page 219: ...c All rights reserved worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Dimensional data is based on nominal values Products may be subject to regional market cons...

Page 220: ...ll rights reserved worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Dimensional data is based on nominal values Products may be subject to regional market conside...

Page 221: ...All rights reserved worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Dimensional data is based on nominal values Products may be subject to regional market consid...

Page 222: ... All rights reserved worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Dimensional data is based on nominal values Products may be subject to regional market consi...

Page 223: ... worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Dimensional data is based on nominal values Products may be subject to regional market considerations License pr...

Page 224: ...d worldwide No part of this drawing may be reproduced stored in a retrieval system or transmitted in any form or by any means electronic mechanical photocopying recording or otherwise without written permission from FLIR Systems Inc Specifications subject to change without further notice Dimensional data is based on nominal values Products may be subject to regional market considerations License p...

Page 225: ...o isolated 0 1 5 V low 3 25 V high 2 IN 2 opto isolated 0 1 5 V low 3 25 V high 3 OUT 1 opto isolated ON supply max 100 mA OFF open 4 OUT 2 opto isolated ON supply max 100 mA OFF open 5 Supply VCC 6 24 VDC max 200 mA 6 Supply Gnd Gnd Note Cables for digital I O ports should be 100 m 328 maximum T559954 r AL 37426 37426 en US 211 ...

Page 226: ...Digital I O connection diagram 29 T559954 r AL 37426 37426 en US 212 ...

Page 227: ......

Page 228: ...CE Declaration of conformity 30 T559954 r AL 37426 37426 en US 214 ...

Page 229: ......

Page 230: ...mmercial lens cleaning liquid with more than 30 isopropyl alcohol 96 ethyl alcohol C2H5OH 31 2 2 Equipment Cotton wool 31 2 3 Procedure Follow this procedure 1 Soak the cotton wool in the liquid 2 Twist the cotton wool to remove excess liquid 3 Clean the lens one time only and discard the cotton wool WARNING Make sure that you read all applicable MSDS Material Safety Data Sheets and warning labels...

Page 231: ...the dust cannot be removed by following this procedure the infrared de tector must be cleaned mechanically This mechanical cleaning must be carried out by an authorized service partner CAUTION In Step 2 below do not use pressurized air from pneumatic air circuits in a workshop etc as this air usu ally contains oil mist to lubricate pneumatic tools 31 3 2 Procedure Follow this procedure 1 Remove th...

Page 232: ...r water damage 32 1 2 Figure The image below shows extensive water damage on an external wall where the water has penetrated the outer facing because of an incorrectly installed window ledge 32 2 Faulty contact in socket 32 2 1 General Depending on the type of connection a socket has an improperly connected wire can re sult in local temperature increase This temperature increase is caused by the r...

Page 233: ...rared image as local temperature increase A socket s construction may differ dramatically from one manufacturer to another For this reason different faults in a socket can lead to the same typical appearance in an infrared image Local temperature increase can also result from improper contact between a wire and socket or from difference in load 32 3 2 Figure The image below shows a series of fuses...

Page 234: ...e area where air is penetrating the frame of the building When you are inspecting a building the temperature difference between the inside and outside should be at least 10 C 18 F Studs water pipes concrete columns and similar components may resemble an insulation deficiency in an infrared image Minor differences may also occur naturally 32 4 2 Figure In the image below insulation in the roof fram...

Page 235: ...ld be sub atmospheric pressure in the house Close all doors windows and ventilation ducts and allow the kitchen fan to run for a while before you take the infrared images An infrared image of draft often shows a typical stream pattern You can see this stream pattern clearly in the picture below Also keep in mind that drafts can be concealed by heat from floor heating circuits 32 5 2 Figure The ima...

Page 236: ...Application examples 32 T559954 r AL 37426 37426 en US 222 ...

Page 237: ...uired several companies with world leading expertise in sensor technologies Extech Instruments 2007 Ifara Tecnologías 2008 Salvador Imaging 2009 OmniTech Partners 2009 Directed Perception 2009 Raymarine 2010 ICx Technologies 2010 TackTick Marine Digital Instruments 2011 Aerius Photonics 2011 Lorex Technology 2012 Traficon 2012 MARSS 2013 DigitalOptics micro optics business 2013 DVTEL 2015 Figure 3...

Page 238: ...lectronics to final testing and calibration all production steps are carried out and supervised by our own engineers The in depth expertise of these infrared specialists ensures the accuracy and reliability of all vital components that are assembled into your infrared camera 33 1 More than just an infrared camera At FLIR Systems we recognize that our job is to go beyond just producing the best inf...

Page 239: ...ms operates a worldwide service network to keep your camera running at all times If you discover a problem with your camera local service centers have all the equip ment and expertise to solve it within the shortest possible time Therefore there is no need to send your camera to the other side of the world or to talk to someone who does not speak your language T559954 r AL 37426 37426 en US 225 ...

Page 240: ... adjust A function that adjusts the image The function works all the time continuously adjusting brightness and contrast according to the im age content convection Convection is a heat transfer mode where a fluid is brought into mo tion either by gravity or another force thereby transferring heat from one place to another dual isotherm An isotherm with two color bands instead of one emissivity emi...

Page 241: ...NETD Noise equivalent temperature difference A measure of the image noise level of an IR camera noise Undesired small disturbance in the infrared image object parameters A set of values describing the circumstances under which the meas urement of an object was made and the object itself such as emis sivity reflected apparent temperature distance etc object signal A non calibrated value related to ...

Page 242: ...ich is the result of a subtraction between two temperature values temperature range The current overall temperature measurement limitation of an IR cam era Cameras can have several ranges Expressed as two blackbody temperatures that limit the current calibration temperature scale The way in which an IR image currently is displayed Expressed as two temperature values limiting the colors thermogram ...

Page 243: ...object and the camera The relative humidity Temperature of the atmosphere 35 2 Emissivity The most important object parameter to set correctly is the emissivity which in short is a measure of how much radiation is emitted from the object compared to that from a perfect blackbody of the same temperature Normally object materials and surface treatments exhibit emissivity ranging from approxi mately ...

Page 244: ...Look for possible reflection sources considering that the incident angle reflection an gle a b Figure 35 1 1 Reflection source 2 If the reflection source is a spot source modify the source by obstructing it using a piece if cardboard Figure 35 2 1 Reflection source T559954 r AL 37426 37426 en US 230 ...

Page 245: ...ed for two important reasons A thermocouple does not measure radiation intensity A thermocouple requires a very good thermal contact to the surface usually by gluing and covering the sensor by a thermal isolator 35 2 1 1 2 Method 2 Reflector method Follow this procedure 1 Crumble up a large piece of aluminum foil 2 Uncrumble the aluminum foil and attach it to a piece of cardboard of the same size ...

Page 246: ... of the tape usually 0 97 8 Measure the temperature of the tape using one of the following measurement functions Isotherm helps you to determine both the temperature and how evenly you have heated the sample Spot simpler Box Avg good for surfaces with varying emissivity 9 Write down the temperature 10 Move your measurement function to the sample surface 11 Change the emissivity setting until you r...

Page 247: ...lf is detected by the camera 35 5 Relative humidity The camera can also compensate for the fact that the transmittance is also dependent on the relative humidity of the atmosphere To do this set the relative humidity to the correct value For short distances and normal humidity the relative humidity can normally be left at a default value of 50 35 6 Other parameters In addition some cameras and ana...

Page 248: ...p a systematic experiment with the objective of finding a single material that would give the desired reduction in brightness as well as the maximum reduction in heat He began the experiment by actually repeating Newton s prism experiment but looking for the heating effect rather than the visual distri bution of intensity in the spectrum He first blackened the bulb of a sensitive mercury in glass ...

Page 249: ...to be true only until 1830 when the Italian investigator Melloni made his great discovery that naturally occurring rock salt NaCl which was available in large enough natural crystals to be made into lenses and prisms is remark ably transparent to the infrared The result was that rock salt became the principal infrared optical material and remained so for the next hundred years until the art of syn...

Page 250: ...uidance An infrared search system tested during this period was able to detect an approaching airplane at a distance of 1 5 km 0 94 miles or a person more than 300 me ters 984 ft away The most sensitive systems up to this time were all based upon variations of the bolometer idea but the period between the two wars saw the development of two revolutionary new infrared detectors the image converter ...

Page 251: ...the limit of visual perception in the deep red At the long wavelength end it merges with the microwave radio wavelengths in the millimeter range The infrared band is often further subdivided into four smaller bands the boundaries of which are also arbitrarily chosen They include the near infrared 0 75 3 μm the middle infrared 3 6 μm the far infrared 6 15 μm and the extreme infrared 15 100 μm Altho...

Page 252: ...m temperature generates blackbody radiation the characteristics of which are determined solely by the temperature of the cav ity Such cavity radiators are commonly used as sources of radiation in temperature refer ence standards in the laboratory for calibrating thermographic instruments such as a FLIR Systems camera for example If the temperature of blackbody radiation increases to more than 525 ...

Page 253: ...it approaches zero again at very long wavelengths The higher the temperature the shorter the wavelength at which maximum occurs Figure 37 4 Blackbody spectral radiant emittance according to Planck s law plotted for various absolute temperatures 1 Spectral radiant emittance W cm2 103 μm 2 Wavelength μm 37 3 2 Wien s displacement law By differentiating Planck s formula with respect to λ and finding ...

Page 254: ... K the peak of radiant emittance lies at 9 7 μm in the far infra red while at the temperature of liquid nitrogen 77 K the maximum of the almost insignifi cant amount of radiant emittance occurs at 38 μm in the extreme infrared wavelengths Figure 37 6 Planckian curves plotted on semi log scales from 100 K to 1000 K The dotted line represents the locus of maximum radiant emittance at each temperatur...

Page 255: ...ended wavelength region although they may approach the blackbody behavior in certain spectral intervals For ex ample a certain type of white paint may appear perfectly white in the visible light spec trum but becomes distinctly gray at about 2 μm and beyond 3 μm it is almost black There are three processes which can occur that prevent a real object from acting like a blackbody a fraction of the in...

Page 256: ...h wavelength According to Kirchhoff s law for any material the spectral emissivity and spectral absorp tance of a body are equal at any specified temperature and wavelength That is From this we obtain for an opaque material since αλ ρλ 1 For highly polished materials ελ approaches zero so that for a perfectly reflecting material i e a perfect mirror we have For a graybody radiator the Stefan Boltz...

Page 257: ...some of it is reflected back into the interior The back reflected radiation is again partially absorbed but some of it arrives at the other sur face through which most of it escapes part of it is reflected back again Although the pro gressive reflections become weaker and weaker they must all be added up when the total emittance of the plate is sought When the resulting geometrical series is summe...

Page 258: ... least to a trained operator It is then his responsibility to modify the measurement situation to avoid the disturbance e g by changing the viewing direction shielding off intense radia tion sources etc Accepting the description above we can use the figure below to derive a formula for the calculation of the object temperature from the calibrated camera output Figure 38 1 A schematic representatio...

Page 259: ...ion power can now be written Equation 2 We multiply each term by the constant C of Equation 1 and replace the CW products by the corresponding U according to the same equation and get Equation 3 Solve Equation 3 for Uobj Equation 4 This is the general measurement formula used in all the FLIR Systems thermographic equipment The voltages of the formula are Table 38 1 Voltages Uobj Calculated camera ...

Page 260: ...allowed to use the cal ibration curve above the highest calibration point what we call extrapolation Imagine that we in a certain case measure Utot 4 5 volts The highest calibration point for the camera was in the order of 4 1 volts a value unknown to the operator Thus even if the object hap pened to be a blackbody i e Uobj Utot we are actually performing extrapolation of the calibration curve whe...

Page 261: ...ected radiation Atm atmosphere radia tion Fixed parameters τ 0 88 Trefl 20 C 68 F Tatm 20 C 68 F Figure 38 3 Relative magnitudes of radiation sources under varying measurement conditions LW camera 1 Object temperature 2 Emittance Obj Object radiation Refl Reflected radiation Atm atmosphere radia tion Fixed parameters τ 0 88 Trefl 20 C 68 F Tatm 20 C 68 F T559954 r AL 37426 37426 en US 247 ...

Page 262: ...mote Sensing 8 Kern Evaluation of infrared emission of clouds and ground as measured by weather satellites Defence Documentation Center AD 617 417 9 Öhman Claes Emittansmätningar med AGEMA E Box Teknisk rapport AGEMA 1999 Emittance measurements using AGEMA E Box Technical report AGEMA 1999 10 Matteï S Tang Kwor E Emissivity measurements for Nextel Velvet coating 811 21 between 36 C AND 82 C 11 Loh...

Page 263: ...50 500 T 0 2 0 3 1 Aluminum polished 50 100 T 0 04 0 06 1 Aluminum polished plate 100 T 0 05 4 Aluminum polished sheet 100 T 0 05 2 Aluminum rough surface 20 50 T 0 06 0 07 1 Aluminum roughened 27 10 µm 0 18 3 Aluminum roughened 27 3 µm 0 28 3 Aluminum sheet 4 samples differently scratched 70 SW 0 05 0 08 9 Aluminum sheet 4 samples differently scratched 70 LW 0 03 0 06 9 Aluminum vacuum deposited ...

Page 264: ...8 5 Brick common 17 SW 0 86 0 81 5 Brick Dinas silica glazed rough 1100 T 0 85 1 Brick Dinas silica refractory 1000 T 0 66 1 Brick Dinas silica un glazed rough 1000 T 0 80 1 Brick firebrick 17 SW 0 68 5 Brick fireclay 1000 T 0 75 1 Brick fireclay 1200 T 0 59 1 Brick fireclay 20 T 0 85 1 Brick masonry 35 SW 0 94 7 Brick masonry plastered 20 T 0 94 1 Brick red common 20 T 0 93 2 Brick red rough 20 T...

Page 265: ...0 38 1 Clay fired 70 T 0 91 1 Cloth black 20 T 0 98 1 Concrete 20 T 0 92 2 Concrete dry 36 SW 0 95 7 Concrete rough 17 SW 0 97 5 Concrete walkway 5 LLW 0 974 8 Copper commercial burnished 20 T 0 07 1 Copper electrolytic care fully polished 80 T 0 018 1 Copper electrolytic polished 34 T 0 006 4 Copper molten 1100 1300 T 0 13 0 15 1 Copper oxidized 50 T 0 6 0 7 1 Copper oxidized to blackness T 0 88 ...

Page 266: ...20 LLW 0 849 8 Granite rough 21 LLW 0 879 8 Granite rough 4 different samples 70 SW 0 95 0 97 9 Granite rough 4 different samples 70 LW 0 77 0 87 9 Gypsum 20 T 0 8 0 9 1 Ice See Water Iron and steel cold rolled 70 SW 0 20 9 Iron and steel cold rolled 70 LW 0 09 9 Iron and steel covered with red rust 20 T 0 61 0 85 1 Iron and steel electrolytic 100 T 0 05 4 Iron and steel electrolytic 22 T 0 05 4 I...

Page 267: ...ted heavily 17 SW 0 96 5 Iron and steel rusty red 20 T 0 69 1 Iron and steel shiny oxide layer sheet 20 T 0 82 1 Iron and steel shiny etched 150 T 0 16 1 Iron and steel wrought carefully polished 40 250 T 0 28 1 Iron galvanized heavily oxidized 70 SW 0 64 9 Iron galvanized heavily oxidized 70 LW 0 85 9 Iron galvanized sheet 92 T 0 07 4 Iron galvanized sheet burnished 30 T 0 23 1 Iron galvanized sh...

Page 268: ...acquer black shiny sprayed on iron 20 T 0 87 1 Lacquer heat resistant 100 T 0 92 1 Lacquer white 100 T 0 92 2 Lacquer white 40 100 T 0 8 0 95 1 Lead oxidized at 200 C 200 T 0 63 1 Lead oxidized gray 20 T 0 28 1 Lead oxidized gray 22 T 0 28 4 Lead shiny 250 T 0 08 1 Lead unoxidized polished 100 T 0 05 4 Lead red 100 T 0 93 4 Lead red powder 100 T 0 93 1 Leather tanned T 0 75 0 80 1 Lime T 0 3 0 4 1...

Page 269: ...ron unpolished 20 T 0 11 0 40 1 Nickel electroplated on iron unpolished 22 T 0 11 4 Nickel electroplated polished 20 T 0 05 2 Nickel oxidized 1227 T 0 85 4 Nickel oxidized 200 T 0 37 2 Nickel oxidized 227 T 0 37 4 Nickel oxidized at 600 C 200 600 T 0 37 0 48 1 Nickel polished 122 T 0 045 4 Nickel wire 200 1000 T 0 1 0 2 1 Nickel oxide 1000 1250 T 0 75 0 86 1 Nickel oxide 500 650 T 0 52 0 59 1 Oil ...

Page 270: ... 4 different colors 70 LW 0 92 0 94 9 Paper black T 0 90 1 Paper black dull T 0 94 1 Paper black dull 70 SW 0 86 9 Paper black dull 70 LW 0 89 9 Paper blue dark T 0 84 1 Paper coated with black lacquer T 0 93 1 Paper green T 0 85 1 Paper red T 0 76 1 Paper white 20 T 0 7 0 9 1 Paper white bond 20 T 0 93 2 Paper white 3 different glosses 70 SW 0 76 0 78 9 Paper white 3 different glosses 70 LW 0 88 ...

Page 271: ...6 1 Platinum wire 50 200 T 0 06 0 07 1 Porcelain glazed 20 T 0 92 1 Porcelain white shiny T 0 70 0 75 1 Rubber hard 20 T 0 95 1 Rubber soft gray rough 20 T 0 95 1 Sand T 0 60 1 Sand 20 T 0 90 2 Sandstone polished 19 LLW 0 909 8 Sandstone rough 19 LLW 0 935 8 Silver polished 100 T 0 03 2 Silver pure polished 200 600 T 0 02 0 03 1 Skin human 32 T 0 98 2 Slag boiler 0 100 T 0 97 0 93 1 Slag boiler 14...

Page 272: ... 100 T 0 07 2 Titanium oxidized at 540 C 1000 T 0 60 1 Titanium oxidized at 540 C 200 T 0 40 1 Titanium oxidized at 540 C 500 T 0 50 1 Titanium polished 1000 T 0 36 1 Titanium polished 200 T 0 15 1 Titanium polished 500 T 0 20 1 Tungsten 1500 2200 T 0 24 0 31 1 Tungsten 200 T 0 05 1 Tungsten 600 1000 T 0 1 0 16 1 Tungsten filament 3300 T 0 39 1 Varnish flat 20 SW 0 93 6 Varnish on oak parquet floo...

Page 273: ...ne 4 different samples 70 SW 0 67 0 75 9 Wood pine 4 different samples 70 LW 0 81 0 89 9 Wood planed 20 T 0 8 0 9 1 Wood planed oak 20 T 0 90 2 Wood planed oak 70 SW 0 77 9 Wood planed oak 70 LW 0 88 9 Wood plywood smooth dry 36 SW 0 82 7 Wood plywood untreated 20 SW 0 83 6 Wood white damp 20 T 0 7 0 8 1 Zinc oxidized at 400 C 400 T 0 11 1 Zinc oxidized surface 1000 1200 T 0 50 0 60 1 Zinc polishe...

Page 274: ...2016 01 13 T505824 xml en US 32182 2016 01 13 T505825 xml en US 32182 2016 01 13 T505826 xml en US 32182 2016 01 13 T505827 xml en US 32521 2016 01 19 T505657 xml en US 37264 2016 09 16 T505658 xml en US 37152 2016 09 08 T505659 xml en US 32063 2016 01 08 T505829 xml en US 32182 2016 01 13 T505661 xml en US 32047 2016 01 08 T505662 xml en US 37153 2016 09 08 T505765 xml en US 29249 2015 10 09 T505...

Page 275: ......

Page 276: ... support flir com Copyright 2016 FLIR Systems Inc All rights reserved worldwide Disclaimer Specifications subject to change without further notice Models and accessories subject to regional market considerations License procedures may apply Products described herein may be subject to US Export Regulations Please refer to exportquestions flir com with any questions ...

Reviews: