39
EN
10.0 TECHNICAL DATA
Housing dimensions
196 x 170 x 107 mm (7.7” x 6.6” x 4.2”)
Weight
1.5 kg (3.3 lbs)
Compressor pressure range
30 to 36 psi (2.1 to 2.5 bar)
Compressor working pressure 8 to 16 psi (0.5 to 1.0 bar)
Airflow
6 ~ 8 l/min
Aerosol droplet size
0.5 to 10 μm
Droplet size according to
MMAD
Less than 4 μm
Maximum medication capacity 6 ml
Nebulisation capacity
>0.5 ml/min in the case of continuous inhalation (locking device pressed or secured)
Operating noise
Approx. 60 dBA
Ambient conditions
Transport/storage
Operation
-20 C to 70 C (-4 F to 158 F)
10 to 95 % relative humidity
10 C to 40 C (50 F to 104 F)
30 to 75 % relative humidity
Input voltage
AC 230 V, 50 Hz
Output current
0.8 A
Power admission
180 VA
Classification according to EN
60601-1
Class II (double insulated)
Type BF
Permanent operation
Protection type
IP 21