![FAG Detector III Manual Download Page 108](http://html1.mh-extra.com/html/fag/detector-iii/detector-iii_manual_3173195108.webp)
108
Heading
Contents
GUID
Global unique identification number of the measuring point.
Timesignal count /
FFT count
Number of time signals and FFT values
·
The time signals and FFT values are stored in four columns under the header.
o
Column 1 contains the time of the measurement in seconds from the start of
the time signal, column 2 contains the time signal measurement.
o
Column 3 contains the frequencies of the FFT in Hz, column 4 the amplitudes.
Measuring point from a balancing configuration
If you export a measuring point from a balancing configuration, the Trendline
software saves a general export file with details on the configuration, the balancing
jobs and the weights. In addition, the program creates one file per balancing step
("Data File<n>.csv") containing the associated sensor data.
The general export file contains the following details:
Heading
Contents
Data File
Export file with the sensor data.
Balancing measurepoint
Name of measuring point
Job Number
Number of the balancing job.
App. weight 1 amplitude
Amplitude of the weight used in plane 1.
App. weight 1 angle
Position angle of the weight used in plane 1.
App. weight 2 amplitude
Amplitude of the weight used in plane 2.
App. weight 2 angle
Position angle of the weight used in plane 2.
comment
Measurement comment.
step_type_text
Type of balancing step (reference run, trial run, etc.).
remove_weights_text
Indicates what weights were removed.
The sensor data files contain the following details:
Heading
Contents
Sensor position
Name of sensor position.
Amp. Coeff. plane 1
Amplitude of the coefficients in plane 1.
Ph. Coeff. plane 1
Angle of the coefficients in plane 1.