PAGE 4(TEST & DATA MANIPULATION(2))
Set up the items related with the test and data control which is necessary in the process
of test.
.
[Figure 8-11] TEST & DATA MANIPULATION(2)
①
DEFAULT PROGRAM: Designate basic programs for vision test. Can choose one
among 12 Custom Program incluging STANDARD but the initial value is
STANDARD.
②
FOG: BALANCE: Designate the value of Fogging for the Balance Test of Binocular.
You can choose one from +0.00D to +1.00D but the initial value is +0.00D.
③
FOG: RED/GREEN: Designate the value of Fogging for the Red/Green Test. You
can choose one from +0.00D to +1.00D but the initial value is +0.00D.
④
INPUT FUSION: Determine whether to put in the test value of Worth-4-Dots. You
can choose YES or NO but the initial value is YES.
⑤
INPUT STEREO: Determine whether to put in the test value of Stereo Test. You
can choose YES or NO, but the initial value is YES.
⑥
INPUT ANISEIKONIA: Determine whether to put in the test value of Aniseikonia.
You can choose YES or NO, but the initial value is YES.
⑦
STARTUP MODE: Designate the start mode when the system was initialized by
[CLEAR] button. You can choose one among EMPTY, UA, LM, RK, and SUB, but
the initial value EMPTY.
⑧
SUBJ STARTUP: Designate the mode on which the information is copied, when
you entered SUB mode by pressing [SUB] button. You can choose RK, LM, or
ANY, but the initial value is ANY.
⑨
SHIFT+FIN: Designate the mode when pressing [SHIFT] and [FIN] together. You
can choose one among LM, RK, SUB but the initial vaule is LM.