Negative Relative Accommodation test
Purpose: In subjective test, estimate the negative relative accommodation.
Chart: Near Vision Chart
Auxiliary Lens: None
Target: Calculate the value of BREAK and RECOVERY for left, right and both eyes.
Way of Test
1. On Special Chart clause of touch screen, press [NADD #] and [NRA] buttons
to enter the Negative Relative Accommodation Test Mode or you can setup the
Negative Relative Accommodation Test Mode by pressing [TEST] button and
choosing [NEGATIVE RELATIVE ACCOM.].
2. Press [R] [BIN] [L] to choose the field for test.
3. Turn the dial counterclockwise
.
When the chart is beginning to be blurred, press (RECOVERY) on the left dialog box
of touch screen.
4. If recovered point is found turning the dial clockwise, press [OK] button to conclude
the test on the left dialog box of touch screen.
[Figure 10-5] Negative Relative Accommodation Test