background image

17

CUSTOMER SETTINGS MENU FUNCTIONS
Within the Customer Settings menu there are several functions that provide customer feedback. The functions 
include an optional inspection reminder, lift usage measured in minutes, and diagnostic information. Below is a 
description of each function.

Inspection Reminder – Optional
NOTE:  A battery must be inserted in the unit at all times for the inspection reminder to be accurate.
The Inspection Reminder function tracks the last inspection occurrence. When the duration of time since the 
last inspection exceeds 30 days, a reminder message will appear when the unit is turned on. The purpose of the 
inspection reminder is to notify the user that the EZ Way Smart Lift

®

 needs to be inspected using the EZ Way 

Safety & Maintenance Checklist form included on pages 13 - 14. When the inspection is complete the reminder 
can be reset in the Customer Settings menu as described below:
1)  To enable/disable the Inspection Reminder enter the Customer Settings menu as described on page 13.
2)  Press the right toggle switch to enter. 
3)  Scroll to Inspection Reminder using the UP and DOWN keys on the controls. 
4)  Press the right toggle switch to enter. 
5)  Scroll to choose Yes or No. 
6)  Press the right toggle switch to enter.
7)  Press the left toggle switch twice to exit.

Inspection Complete
If ‘INSPECTION OVERDUE’ appears when the EZ Way Smart Lift

®

 is turned on, inspect the unit following the 

EZ Way Safety & Maintenance Checklist’.  The EZ Way Smart Lift

®

 will continue to operate normally.

Once inspection is complete, enter the Customer Settings Menu as described on page 13.
1)  Press the right toggle switch to enter. 
2)  Scroll using the UP and DOWN keys to Inspection Complete. 
3)  Press the right toggle switch to enter. 
4)  Scroll to choose Yes. 
5)  Press the right toggle switch to enter.
6)  Press the left toggle switch twice to exit.

Overweight Counts – Available on EZ Way Smart Lift

®

 with scale

The Overweight Counts function tracks weight overload occurrences. When a weight overload occurs the 
Overweight Counts register increments by one. 

Actuator Minutes
The Actuator Minutes function tracks the accumulated amount of raising and lowering time of the actuator 
by the minute. With each minute of runtime the Actuator Minutes register increments by one. This read-only 
information provides accurate customer feedback of EZ Way Smart Lift

®

 usage within the facility and can be 

a valuable tool in monitoring a safe patient handling program. In the event of an actuator replacement the 
Actuator Minutes register can be reset.

Actuator Overload Counts   
The Actuator Overload Counts function tracks any actuator overload occurrences. When an actuator overload 
occurs the Actuator Overload Counts register increments by one. Potential causes of an actuator overload 
include accidental lifting of the EZ Way Smart Lift

®

 against an obstruction, lifting a patient that is heavier than 

the rated capacity of the unit, or mechanical failure (internal actuator binding).

Lift/Stand Total Minute Usage Count
The Actuator Total Minutes function tracks the accumulated amount of raising and lowering time of the unit. 
The Lift/Stand Total Minutes is not reset when an actuator is replaced. This read-only information provides 
accurate customer feedback of EZ Way Smart Lift

®

 usage for the entire life of the unit.

Summary of Contents for Smart Lift

Page 1: ...t Lift 500 600 1 000 lb Capacities Form 2 174 Rev 8 10 18 Operator s Instructions EZ Way Inc PO Box 89 Clarinda IA 51632 1 800 627 8940 www ezlifts com WARNING Cancer and Reproductive Harm www P65Warn...

Page 2: ...when lifting or lowering a patient The EZ Way Smart Lift was designed to be operated safely by one caregiver However depending on the situation facility policy and the patient s condition two caregiv...

Page 3: ...nd service will be free from defects in material and workmanship for a period of see above from the date of purchase by the original purchaser Normal wear and tear injury by natural forces user neglec...

Page 4: ...he ON OFF button The EZ Way Smart Lift will display a greeting message while several diagnostic functions are performed The EZ Way Smart Lift has a built in timer that automatically shuts off the unit...

Page 5: ...arious sizes Please contact EZ Way with any questions regarding sling dimensions and sizes or view them at www ezlifts com We have used the Medium Deluxe Sling throughout the manual because it is the...

Page 6: ...patient 1 Make sure there are no cords or other objects near the path of the lift or under or near the bed that could obstruct the wheels of the lift 2 Do not lock the wheels of the EZ Way Smart Lift...

Page 7: ...the actuator motor as a handle for moving the EZ Way Smart Lift WARNING Do not place any objects around the actuator motor or between the actuator and the lift mast Step 6 Lower patient into wheelchai...

Page 8: ...distributed on the sling NOTE Make sure all seams of the sling are smooth underneath the patient Step 2 Position EZ Way Smart Lift 1 Make sure there are no cords or other objects near the path of the...

Page 9: ...6 Lower the patient onto bed 1 Raise or lower the patient to a height necessary to clear the bed Move the patient over the bed 2 Push the DOWN button on the hand control 3 When the patient s weight is...

Page 10: ...objects or beneath the casters or legs of the lift The sling should be free from obstruction Step 4 Follow same procedures used when transferring patient and when lowering patient to bed chair or whe...

Page 11: ...than 11 0 is used rinse twice 5 Tumble dry temperature 100 F maximum High heat will weaken the fabric 6 Snap the buckle together before washing and drying This will prevent any damage to the plastic b...

Page 12: ...he Operating Instructions 3 Raise the patient Be sure no other object is touching the lift 4 Press the weight key The lift will display the weight and when the sign changes to a the weight will be loc...

Page 13: ...between hanger assembly and load cell the hanger assembly needs to be replaced 2 Check all bolts to ensure they are tight 3 Check boom to hanger assembly pivot bolt peel rubber back to assure nut is...

Page 14: ...when lowering the lift boom if an obstruction is encountered 9 Check all the wheels and brakes to make sure they are functioning properly Hair that gets picked up and wrapped in between wheels should...

Page 15: ...be accessed at any time Possible causes of actuator overload include lifting a patient that is heavier than the rated capacity of the unit if the Smart Lift is obstructed by another object during lif...

Page 16: ...ings menu Entering the Customer Settings menu After the unit has been turned on press and hold the DOWN key followed by the UP key holding both keys for 5 seconds To back out of the Customer Settings...

Page 17: ...ettings Menu as described on page 13 1 Press the right toggle switch to enter 2 Scroll using the UP and DOWN keys to Inspection Complete 3 Press the right toggle switch to enter 4 Scroll to choose Yes...

Page 18: ...ce legs of sling have been fitted underneath legs and crossed attach two loops of same color to the hangar bar hooks located nearest the feet Ensure all loops securely fastened to hooks before lifting...

Page 19: ...19...

Page 20: ...Sling Size1 Small Medium Large X Large XX Large XXX Large Color Coding2 Gray Beige Burgundy Green Black Brown Weight of Patient1 70 100 lbs 90 220 lbs 190 320 lbs 280 450 lbs 400 600 lbs 600 lbs Maxim...

Reviews: