15
Customer settings summary sheet
Guide to main functions and keys
Menu
list
Description
Inspection Reminder
Enable/disable, Occurs every 30 days
Inspection Complete
Resets 30-day counter
Overweight Counts
Counts occurrences of lifting beyond a unit’s rated capacity, available on
units with scale
Actuator Minutes
Actuator usage counter, should be reset upon actuator replacement
Actuator Overload Counts
Counts occurrences of actuator overloads, should be reset upon actuator
replacement
Lift/Stand Total Minutes
Stand/Lift usage counter in minutes
Lb./Kg Default Setting
Sets default for weight unit of measure
Navigating through menus key
Scrolling to change menu item
UP/DOWN Keys
Key to enter data
Right Toggle Switch
Key to exit menu
Left Toggle Switch
Main Functions
Key Combination
ON/OFF
Turns unit on and off
Weight Recall
DOWN + Weight
Customer Settings Menu
DOWN + UP – hold 5 seconds
UP
Used to raise the EZ Way Smart Lift
®
boom upward
DOWN
Used to lower the EZ Way Smart Lift
®
boom
Zero
Used to zero out the scale
Weight
Used to take weight readings on units with scale
LB/KG
Switches weight reading between pounds and kilograms
Diagnostic messaging
Actuator Overload
When an actuator overload occurs the EZ Way Smart Lift
®
computer records the event to protect the unit’s
components, and an ‘ACTUATOR OVERLOAD’ message will display. An event count is stored in the
Customer Settings menu under ‘Actuator Overload Counts’. This read-only count can be accessed at any time.
Possible causes of actuator overload include lifting a patient that is heavier than the rated capacity of the unit, if
the Smart Lift is obstructed by another object during lifting or mechanical failure occurs (internal actuator
binding, etc.). In the event of an actuator overload, the EZ Way Smart Lift
®
will only operate in the DOWN
direction.
When the actuator overload count is greater than 4, a fault message will display when the unit is turned on,
reading ‘ACTUATOR OVERLOAD…CALL EZ WAY SERVICE’. When this message appears call the EZ Way
Service Department at 1-800-627-8940.
Weight Overload – Available for EZ Way Smart Lift
®
with Scale
When an UP or DOWN command is executed the circuit board monitors weight readings from the scale load
cell. If the weight measured is higher than the rated capacity of the EZ Way Smart Lift
®
the control will
recognize this condition and a ‘WEIGHT OVERLOAD’ message will display. An event count is stored in the
Customer Settings menu under ‘Weight Overload Counts’. This read-only count can be accessed at any time.
The weight overload may result from lifting a patient that is heavier than the rated capacity of the unit, if the
Smart Lift is obstructed by another object during lifting or a scale load cell failure occurs.
To override a ‘WEIGHT OVERLOAD’ press the DOWN key. The EZ Way Smart Lift
®
will only operate in
the DOWN direction during a weight overload. When the weight overload count is greater than 4 a fault
message will display when the unit is turned on. The message reads ‘WEIGHT OVERLOAD…CALL EZ WAY
SERVICE’. When this message appears call the EZ Way Service Department at 1-800-627-8940.
Summary of Contents for Smart Lift