- 15 –
8. Adjusting the position of the glass trap ball
(
Adjusting the manual slide
)
Use both hands to lift the manual
slide slowly.
After loosening the manual slide knob bolt, the
drive body may fall down, glass may break and
result in a personal injury if you try to lift the body
suddenly.
Caution
You can adjust the manual slide assembly to a
position appropriate for the trap ball or specimen
flask sizes.
Follow the procedures below to install a trap ball
before installing the cooler.
7. Connecting the power cord
(1) Insert the power cord plug into the inlet on the
back of the stand base.
(2) Make sure that the power switch is turned OFF
and then insert the power plug into an outlet.
Be sure to lower the manual slide
slowly while supporting the driving
assembly with a hand.
When you loosen the manual slide knob bolt
suddenly, the drive may suddenly lower from the
weight of the cooler or sample flask causing the
glass to break from the shock and you might get
injured by the broken glass.
Caution
(1) Loosen the manual slide knob bolt, hold the drive
body with both hands and slowly lift it and then
tighten the manual slide knob bolt.
※
Remove the cooler before attempting these
procedures.
※
The internal tightening plate may come if you
loosen the manual slide knob excessively. If it
occurs, lower the slide body, align the knob
bolt with the internal plate and then tighten the
knob.
(2) Install the trap ball and the specimen flask using
the Eyela clamps.
(3) Set the bath, support the lower part of the front of
the drive with one hand, gently loosen the manual
slide knob bolt to adjust the height.
※
When you loosen the manual slide knob bolt
suddenly, the drive may suddenly lower from
the weight of itself causing the glass from
shocks and you might be injured with broken
glass.
Plug
Inlet
LOCK
FREE
Manual slide knob bolt
Drive body
Slide body
Trap bulb
Sample flask
Eyela clip
Eyela clip