- 14 –
5. Adjusting the angle of the drive body
Adjust the angle of the drive body
while holding the condenser with a
hand.
When you loosen the angle adjusting knob bolt, the
drive may suddenly incline from the weight of the
cooler or sample flask and fall down and you might
get injured by the broken glass.
Caution
(1) Lift the jack body.
(2) Loosen the angle adjusting knob bolt to adjust the
angle while supporting the condenser with a hand.
(3) Check the vertical position of the jack and its
position relative to the bath and then securely
tighten the angle adjusting knob bolt.
6. Connecting the vacuum hose and
the coolant hose
(1) Insert the hose holder into the groove on the right
side of the jack body.
(2) Connect the condenser vacuum nozzle and the
suction port of your vacuum unit using a vacuum
hose through the hose holder.
※
The vacuum hose for connecting to the vacuum
unit is not included. Check the size of the
connection nozzle and prepare a correct hose.
※
Take care not to apply excessive force when
connecting the vacuum hose. The nozzle is
made of resin and may be damaged if it is
subject to an excessive force.
(3) Connect the cooler coolant nozzle and the
circulation port of you circulating chiller using a
connecting hose through the hose holder.
※
The connecting hose is not included. Check the
capacity of your circulation unit and the size of
the connection nozzle and prepare a correct
hose.
Vacuum nozzle
(
white
)
Cooling nozzle inlet
Cooling nozzle outlet
Vacuum nozzle
Cooling nozzle inlet
Cooling nozzle outlet
Vacuum nozzle
LOCK
FREE
Angle adjusting knob bolt
Hose holder
Right side of
the jack body