TKG250 Ultrasonic Thickness Gauge
31
Using the Diff Option
From the
Measurement
screen, there are two Diff options, which include:
Diff ABS:
Absolute where the gauge will display a thickness value as an absolute
number of what has been entered. For example, diff abs = 0.500” and the real
thickness is 1.000 inch, the display will show diff abs 0.500. If the real value is
0.300, the gauge will display, -0.200”
Diff %:
When this mode is on, the displayed value will be listed as a percentage
of the value entered into as the diff. For Example, if the operator set the diff to
percentage at 0.500”, a reading of 0.250” will result in 50%
Using the Alarm Option
From the
Measurement
Screen, the TKG250 thickness gauge offers various alarm
settings to alert you of low or high thresholds via audible sounds, display flashes,
sensational vibrations (if on) and illuminating keypad. There are several; different alarms
options, which include:
Alarm Off
:
Sets the audible alarm either off.
Low Alarm
:
Audible alarm will occur and display will flash when the displayed
reading is less than the low alarm reference value.
High Alarm
:
Audible alarm will occur when the displayed reading is higher than
the high alarm reference value.
High_Low Alarm:
Audible alarm will occur and display will flash when the
displayed reading is less than the low alarm or greater than the High alarm
reference value.
Low Alarm/Vibra
:
Vibration alarm will occur and display will flash when the
displayed reading is less than the low alarm reference value.
High Alarm/Vibra
:
Vibration alarm will occur and display will flash when the
displayed reading is higher than the high alarm reference value.
High_Low Alarm/Vibra:
Audible alarm will occur and display will flash as well
as vibrate when the displayed reading is less than the low alarm or greater than
the High alarm reference value.
High_Low Alarm/Vibra with Dynamic Waveform Adjust:
Audible alarm will
occur and display will flash as well as vibrate when the displayed reading is less
than the low alarm or greater than the High alarm reference value. In addition, the
color of the A-Scan will dynamically change to red, yellow (caution…within an
operators preset value of high or low alarm, i.e. 20%) or green corresponding to
bad, caution or good readings.