background image

475055-EN v2.5   3/14 

3

Operation 

Preparation for Measurement

 

1. 

The 475055 automatically determines Tension or Compression (pull/push) force during use. 
Compression displays as a negative value and Tension displays as a positive value. 

2. 

Select units of measure (kilograms, pounds, or Newtons) via the units select switch. 

3. 

Screw in the sensor hook and loop on the load sensor and then connect the sensor to the 
meter via the input jack on top of the meter. 

4. 

Select FAST (0.2 seconds) or SLOW (0.6 seconds) update rate. 

5. 

Zero the display before each measurement via the Zero button. 
 

Note: 

The sensing head with adapter must be in line with the object being measured. Avoid 
rotating the sensing head. Refer to the Figure 1. 

 
 
 

Figure 1 

 Correct and Incorrect Angles of Measurement 

 

 

 
 
 
 
 
 
 
 
Measurement Mode 

1. 

Slide the POWER switch to the ON (I)position. Reverse the LCD display if desired via the 
Display Reverse key. 

2. 

Zero the meter before each measurement. 

3. 

Touch the adapter to the object being measured in a straight line. Refer to Fig.1. 

4. 

Begin measurement by applying force (Push or Pull). Read the LCD display. 

5. 

After completing the measurement, the display will indicate "0.00" if the position and angle of 
the Force Gauge have not changed. 

Note:

 Be sure to press "ZERO" before taking a new measurement. 

 

Overrange Indication 

The Model 475055 uses a strain gauge sensor to measure force. Avoid subjecting the instrument to 
forces outside the specified measurement range. The sensitive strain gauge can be damaged or 
otherwise compromised by misuse. When an over-range condition is sensed by the meter the LCD 
displays all dashes. 
Note that the dashes are shown at the top of the LCD (

--------

) for tension measurements and at the 

bottom of the display (

--------

) for compression measurements. 

Acceptable

Not Acceptable

Summary of Contents for 475055

Page 1: ...User s Guide Models 475055 Digital Force Gauge ...

Page 2: ...r reverse display direction and a separate sensor An RS 232 PC interface provides data acquisition functionality This meter is shipped fully tested and calibrated and with proper use will provide years of reliable service Meter Description 1 Sensor socket 2 LCD display 3 Keypad 4 Kg N lb switch 5 Off On Peak switch 6 Sensor plug 7 Sensor loop 8 S type load sensor 9 Sensor hook 10 Mounting holes 60...

Page 3: ... Slide the POWER switch to the ON I position Reverse the LCD display if desired via the Display Reverse key 2 Zero the meter before each measurement 3 Touch the adapter to the object being measured in a straight line Refer to Fig 1 4 Begin measurement by applying force Push or Pull Read the LCD display 5 After completing the measurement the display will indicate 0 00 if the position and angle of t...

Page 4: ...ection Press the FAST SLOW key to select the desired display update rate The FAST icon will appear in the display when FAST is selected Mounting For best results mount the Digital Force Gauge to a test fixture Mounting holes are provided on the rear of the meter for use with fixtures and test stands Display Backlight Pressing the backlight button momentarily will turn on the backlight After approx...

Page 5: ... 2 seconds Slow mode 0 6 seconds Over range Indicators LCD displays for tension measurements and for compression measurements Zero Tare Control Up to Maximum capacity Full Scale Deflection 1mm Transducer type S type load cell with 2 hooks thread size M6 x 1mm ISO and 6 5 ft 2m cable Peak Hold Freezes Max reading on display Data Output RS 232 serial interface with 16 bit data stream output Operatin...

Page 6: ...or consequential damages FLIR s total liability is limited to repair or replacement of the product The warranty set forth above is inclusive and no other warranty whether written or oral is expressed or implied Calibration Repair and Customer Care Services FLIR Systems Inc offers repair and calibration services for the Extech Instruments products we sell NIST certification for most products is als...

Page 7: ...tel La responsabilité totale de FLIR est limitée à la réparation ou au remplacement du produit La garantie définie ci dessus est inclusive et aucune autre garantie écrite ou orale n est exprimée ou implicite Calibrage réparation et services après vente FLIR Systems Inc offre des services de calibrage et de réparation pour les produits Extech Instruments que nous commercialisons Nous fournissons ég...

Page 8: ...identales o consecuentes La responsabilidad total de FLIR está limitada a la reparación o reemplazo del producto La garantía precedente es inclusiva y no hay otra garantía ya sea escrita u oral expresa o implícita Servicios de calibración reparación y atención a clientes FLIR Systems Inc ofrece servicios de reparación y calibración para los productos que vendemos de Extech Instruments Además ofrec...

Reviews: