Page 41 of 50
7.5 Environmental Conditions
Operating Environment Conditions
Installation Category II
Pollution Degree 2
Ambient Temperature: 10 to 35 degrees Celsius
Relative Humidity:
15% to 95% (non-condensing)
Atmospheric Pressure: 700 to 1060 hPa
Altitude:
2000 meters (maximum)
Transport and Storage Conditions
Temperature:
-10 to 60 degrees Celsius
Relative Humidity:
10% to 100% (non-
condensing)
Atmospheric Pressure: 500 to 1060 hPa
7.6 Regulatory Compliance Safety:
The R2000 Radiometer has been designed to meet the
requirements of the following standards:
IEC 61010-1:2001/ EN 61010-1:2001
Safety Requirements for Electrical Equipment for
Measurement, Control and Laboratory Use
CAN/ CSA C22.2 No. 61010-1-04
Safety Requirements for Electrical Equipment for
Measurement, Control, and Laboratory Use Part 1;
General Requirements
UL 61010-1: 2
nd
Edition
Electrical Equipment for Laboratory Use Part 1: General
Requirements