Page 18 of 50
downloaded into a Data Log as seen on the R2000
Control Panel (via the GUI software provided). The
stored readings can only be viewed by downloading to a
PC. Once a reading has been stored, it can not be
viewed on the R2000 Radiometer display.
4.12
Interfacing with Compatible OmniCure UV
Curing Systems
Refer to the OmniCure Curing System User Guide.
The R2000 Radiometer is equipped with two I/O ports
for communication with compatible OmniCure curing
systems. When connected, the R2000 Radiometer is
able to calibrate the OmniCure SERIES 2000 UV Curing
System and set the irradiance to a specific level.
To interface via the IR
Window:
Point the IR window located on
the R2000 Radiometer directly
at the OmniCure UV Curing
System ensuring that the
OmniCure is turned ON.
Always keep the R2000
Radiometer pointed at the
OmniCure UV Curing System
while performing a function
that is communicating with the
OmniCure.
The illustration shown
indicates the IRDA field of
view and operating range for
the R2000/ S2000
combination.