Alberta E-Vent User Manual Version 1.3 June 3, 2020
40
FiO2 Titration Look up Table
Table F-3 FiO2 Titration Look up Table
Tidal Volume
Setting
1
Lpm
2
Lpm
3
Lpm
4
Lpm
5
Lpm
6
Lpm
7
Lpm
8
Lpm
9
Lpm
10
Lpm
15
Lpm
45%
49%
73%
86%
98%
99%
100% 100% 100% 100% 100% 100%
50%
46%
65%
78%
97%
98%
99%
99%
99%
99%
99%
99%
55%
43%
59%
71%
94%
97%
98%
98%
98%
99%
99%
99%
60%
40%
55%
66%
84%
93%
97%
98%
98%
98%
99%
99%
65%
38%
50%
61%
75%
86%
96%
97%
98%
98%
98%
99%
70%
36%
47%
57%
69%
80%
92%
96%
98%
98%
98%
99%
75%
35%
44%
53%
64%
74%
86%
95%
97%
97%
98%
99%
80%
34%
42%
50%
60%
70%
81%
88%
96%
97%
97%
98%
85%
32%
40%
48%
57%
66%
76%
84%
93%
96%
97%
98%
90%
32%
39%
46%
54%
62%
72%
79%
87%
95%
96%
98%
95%
31%
38%
44%
52%
59%
68%
75%
83%
92%
95%
98%
100%
30%
36%
43%
50%
57%
64%
71%
79%
86%
93%
98%
NOTE:
FiO2 Titration Look up Table (Table F-3)
is only valid for Ambu-Bag Spur II REF
520211000 with a RR of 16 bpm.