7
Sensafe.com
FILL, DIP, READ
TOTAL IRON
REMOVE REAGENT STRIP
Remove one eXact
®
Strip Micro FE TPTZ and set in a dry,
convenient place. Replace cap on bottle.
5
ZERO METER
Press
and the LCD display reads 0.00PPM,
indicating the meter is ready for testing.
6
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously DIP the eXact
®
Strip by submerging all
pads in the sample then use a gentle, constant, back
and forth motion (2 strokes per second) until the timer
displays '1'. Remove and discard the strip. The display
will automatically start to count up for 40 seconds.
7
READ RESULTS
READ
result displayed as Total Iron. The result is
automatically stored in the FE3 menu. After testing is
complete, rinse the sample cell immediately.
8