15
Sensafe.com
TROUBLESHOOTING
Listed below are possible situations that may arise while testing.
Please contact one of our knowledgeable customer service
representatives if you require further assistance.
Subject
Cause
Solution
Dim screen or no
response from meter
Low battery
Replace batteries
“LO” on LCD while
zeroing
Low battery
Replace batteries
Dirty cell
Clean cell
Cloudy sample
Dilute sample or use
fi
lter
Bad LED
Contact ITS
“HI” on LCD while
reading
Result above
detection level
Re-run test to verify result
“LO” on LCD while
reading
Result below
detection level
Re-run test to verify result
WELL EXACT
®
EZ ACCURACY
All tests have been calibrated using certi
fi
ed reference standards
and analytical spectrophotometric methods. The Well eXact
®
EZ has been factory calibrated and will stay valid because of its
exceptional quality. We are so con
fi
dent in the Well eXact
®
EZ, we
offer an industry leading 2-year warranty.
We built the Well eXact
®
EZ to be easy, accurate and
environmentally friendly. We have achieved this by utilizing our
patented eXact
®
Strip Micro Technology, which uses 60% less
water and chemistry than alternative methods. Instead of using
a 10mL water sample, eXact
®
Strip Micro uses a 4mL water
sample. The accuracy of the meter is maintained by designing
the sample cell with an 11mm path-length.
Ready Snap
®
3 is a calibration veri
fi
cation
solution with predetermined color to verify the
accuracy of your Well eXact
®
EZ photometer.
The easy 3 step procedure (snap,
fi
ll, and
test) allows for quick veri
fi
cation of test
parameters. Each box contains 10 ampoules
of 10mL solution with no dilution necessary.
METHOD VERIFICATION
WELL EXACT
®
EZ ASSIGNED VALUE FOR READY SNAP
®
3
READY SNAP
®
3
TOTAL HARDNESS
DESIRED VALUE
TOTAL HARDNESS
ACCEPTABLE RANGE
Red dye # 505
426 ppm
358 - 494
Note:
Values indicated are speci
fi
c to Well eXact
®
EZ readings and current
concentrations as found at time of manufacture.