8
©
2018, Industrial Test Systems, Inc. Printed in USA
1
PREPARE SAMPLE
Prepare sample with
Mini Dilution Kit II #487202
.
1.
Rinse 50mL graduated tube with salt-free water.
2.
Rinse the 3.0mL syringe with saltwater sample.
3.
Fill the syringe to the 2.0mL line, very precisely, and
add to the graduated tube.
4.
Fill the graduated tube to the 40mL line with salt-free
water. Cap graduated conical tube.
5.
Mix content of graduated tube by turning up side
down at least three times. 1:20 Dilution Sample is
ready for testing.
2
POWER ON PHOTOMETER
Press the
button to power on the Pool eXact
®
EZ.
3
FILL CELL
Before testing, rinse CELL and clean with brush
thoroughly. Finally rinse the cell 3 times with the water
sample to be tested, then
FILL
cell to capacity to
begin test.
4
SELECT TEST
Press and re-press the
button until the display
shows CHH.
5
CAP CELL AND ZERO METER
Place the Cell Cover onto the CELL and press
and the photometer display reads 0
PPM
, indicating the
meter is ready for testing.
CHLORIDE (Salt)
FILL, DIP, READ