REMOVE STRIP

 

Remove one eXact

®

 Strip Micro AL (Part No. 486641) 

and set in a dry, convenient place. Replace cap.

6

DIP STRIP AND PRESS READ 

Press 

 to initiate a 20 second countdown and 

simultaneously 

DIP

 the eXact

®

 strip in the sample, 

gently touching the bottom of the cell. Use a gentle 

constant back and forth motion (2 strokes per 

second) until the timer displays “1”. Be careful to not 

spill the sample from the CELL. Remove and discard 

the strip.

                    

TIP

For water temperatures above 95°F/35°C (hot tubs), Press READ 
when the timer displays “10”, or use the specially calibrated SPA 
eXact

®

 EZ photometer.

7

CAP CELL AND READ RESULTS

Place the Cell Cover onto the CELL and 

READ

 

result displayed as Total Alkalinity. This result is 
automatically stored in the AL(1) menu. After testing 
is complete, rinse the sample cell immediately and 
clean with brush.

      

TIP

If result reads “HI”, repeat steps 5-7 with a Total Alkalinity Range 
Extender Strip (part no. 486665) until a numerical value is displayed; 
Count how many AL Range Extender strips were used; each strip 
adds 130 ppm to the 

fi

 nal result.

If result reads “HI”, repeat steps 5-7 with a Total Alkalinity Range 
Extender Strip (part no. 486665) until a numerical value is displayed; 
Count how many AL Range Extender strips were used; each strip 
adds 130 ppm to the 

fi

 nal result.

TOTAL ALKALINITY

FILL, DIP, READ

Summary of Contents for 486201

Page 1: ...1 exactmeters com QUICK START GUIDE Pool ADVANCED PHOTOMETER SYSTEM DIGITAL ACCURACY FOR BETTER POOL CARE Certified to NSF ANSI Standard 50...

Page 2: ...r Note This system has been calibrated for use with only our eXact Micro reagents CONTENTS Welcome 2 Pool eXact EZ Overview 3 Install AAA Batteries 3 Fill Dip Read Total Alkalinity 4 5 Calcium Hardnes...

Page 3: ...e meter on and creates a baseline for your water testing CELL COVER Covers the cell for mixing and bright light situations Protected against the effects of immersion in water to depth between 15cm and...

Page 4: ...sh thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the button until the display shows AL 1 4 CAP...

Page 5: ...calibrated SPA eXact EZ photometer 7 CAP CELL AND READ RESULTS Place the Cell Cover onto the CELL and READ result displayed as Total Alkalinity This result is automatically stored in the AL 1 menu Aft...

Page 6: ...sh thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the button until the display shows CA 5 4 CAP...

Page 7: ...motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip See pg 23 for important tips TIP For water temperatures above 95 F...

Page 8: ...f graduated tube by turning up side down at least three times 1 20 Dilution Sample is ready for testing 2 POWER ON PHOTOMETER Press the button to power on the Pool eXact EZ 3 FILL CELL Before testing...

Page 9: ...r second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip See pg 23 for important tips 8 CAP CELL AND READ RESULTS Place the Cell Cover onto the...

Page 10: ...testing rinse CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the butto...

Page 11: ...two 2 drops of bleach 5 8 and clean with brush until stain is removed Caution Avoid contact of bleach with eyes and clothing TIP For water temperatures above 95 F 35 C hot tubs press READ when the ti...

Page 12: ...or Free Chlorine 6ppm then LCD will display 0 00ppm indicating the meter is ready for testing 9 REMOVE STRIP Remove one eXact Strip Micro CL DPD 3 Part No 486638 and set in a dry convenient place Repl...

Page 13: ...only Monochloramine interferences are known to occur in free chlorine DPD methods This interference is dependent on temperature and monochloramine concentration Ozone O3 Color similar to other oxidiz...

Page 14: ...e CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and re press the button until the...

Page 15: ...bottom of the cell Use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip The displa...

Page 16: ...2 FILL CELL Before testing rinse CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell to capacity to begin test 3 SELECT TEST Press and...

Page 17: ...Place the cell cover onto the sample cell Press to initiate a 20 second countdown Place thumb or finger over cap and mix the sample by turning the meter upside down repetitively during the 20 second c...

Page 18: ...a ua ua ua ua tua a a a a tu tu ua tua t t t t t t tu u u u u ua ua a a a a a a iiiii ti ti t t ti t ti t tii t t t t tio ti tio i tio t t t tio ti tio o tio tio t t ti tio tio tio t tio o o tio tio t...

Page 19: ...nt back and forth motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip TIP For water temperatures above 95 F 35 C hot tu...

Page 20: ...e filling the meter with the sample to be tested 2 FILL CELL Before testing rinse CELL and clean with brush thoroughly Finally rinse the cell 3 times with the water sample to be tested then FILL cell...

Page 21: ...cell Use a gentle constant back and forth motion 2 strokes per second until the timer displays 1 Be careful to not spill the sample from the CELL Remove and discard the strip The display will automat...

Page 22: ...d if necessary The Acrylic Calibration Key is a tool for verifying the calibration of your Pool eXact EZ Photometer compared to its original factory settings Run this test immediately upon receipt and...

Page 23: ...th water and the brush provided Test immediately after filling the cell with the water sample Due to the strip slitting process you may find one or two strips that are noticeably smaller or larger in...

Page 24: ...ical information and how to videos For additional technical support call 803 329 9712 Industrial Test Systems Inc 1875 Langston Street Rock Hill SC 29730 800 861 9712 or 803 329 9712 its sensafe com s...

Page 25: ...afe com to describe the problem and obtain a return authorization form before returning the photometer to ITS Damage caused by improper packing of the photometer for return shipment to ITS will not be...

Page 26: ...hlorine and the eXact Strip Micro CL DPD 3 for Combined Chlorine and the eXact Strip Micro CL DPD 4 for Total Chlorine meet your reportable testing requirements because the eXact Strip Micro strips de...

Page 27: ...healthy The MAHC requires NSF ANSI 50 certification of water quality testing devices WQTD used in recreational facilities such as public swimming pools interactive fountains and Waterparks Third party...

Page 28: ...25 Chlorine Free DPD 1 2 486637 0 00 12 0 NSF 50 L1 100 Chlorine Combined DPD 3 486638 0 00 12 0 NSF 50 L2 100 Chlorine Total DPD 4 2 486670 0 00 12 0 5 100 Copper 486632 0 00 9 0 2 50 Cyanuric Acid 4...

Reviews: