21
exactmeters.com
PHOSPHATE
FILL, DIP, READ
5
REMOVE STRIP
Remove one eXact
®
Strip Micro PO
4
(Part No. 486814)
and set in a dry, convenient place. Replace cap.
6
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
strip in the sample,
gently touching the bottom of the cell. Use a gentle
constant back and forth motion (2 strokes per
second) until the timer displays
“1”
. Be careful to not
spill the sample from the CELL. Remove and discard
the strip. The display will automatically start to count
up for 120 seconds. See
pg 23
for important tips.
7
CAP CELL AND READ RESULTS
Place the Cell Cover onto the CELL and
READ
result displayed as Phosphate. This result is
automatically stored in the PO4 menu. After testing
is complete, rinse the sample cell immediately and
clean with brush.