19
exactmeters.com
PH
FILL, DIP, READ
5
REMOVE STRIP
Remove one eXact
®
Strip Micro PH (Part No.
486639-II) and set in a dry, convenient place.
Replace cap.
6
DIP STRIP AND PRESS READ
Press
to initiate a 20 second countdown and
simultaneously
DIP
the eXact
®
strip in the sample,
gently touching the bottom of the cell. Use a gentle
constant back and forth motion (2 strokes per
second) until the timer displays
“1”
. Be careful to not
spill the sample from the CELL. Remove and discard
the strip.
TIP
For water temperatures above 95°F/35°C (hot tubs), Press READ
when the timer displays “10”, or use the specially calibrated SPA
eXact
®
EZ photometer.
7
CAP CELL AND READ RESULTS
Place the Cell Cover onto the CELL and
READ
result
displayed as pH. This result is automatically stored in
the PH(2) menu. After testing is complete, rinse the
sample cell immediately and clean with brush.