15
Preset alarms/alerts
The following alarms/alerts are factory preset:
Alarm
Factory Setting
Icon
Temperature Range Alarm
Min 2
°
C Max 8
°
C
Door Ajar Alarm
1 minute
No Power Alarm
Instantaneous
Battery Low Alarm
< 30% Remaining
Memory Full Alarm
< 30% Remaining
No Wifi Alarm
>60 seconds disconnected
Set Point Temperature
When you first turn on the unit, the fan
will run and the refrigerator will work to
reach the default set point temperature of
4 °C. Once set point is reached, the fan
speed will reduce to conserve energy. The
unit will reach the air temperature setting
rapidly. Reaching glycol set point
temperature will take longer.
When the unit is first plugged in and powered, the Temperature Range Alarms are not
enabled until the unit reaches its preset temperature range + an offset of 3 °C. If the set
point temperature is 4 °C, then the alarms are enabled when the unit reaches 7 °C.
Loading Products for the First Time
After you power on your refrigerator, monitor the temperature display on the control panel
and do not load until the default set point of 4 °C (39 °F) is reached. There are three
temperature monitors inside the refrigerator, the glycol sensor and two air sensors. When you