Checking the Glycol Sample Bottle
The glycol sample bottle is pre-installed inside your unit. The sample bottle
allows you to monitor the internal temperature of items being stored in the
refrigerator.
Please verify:
1.
The cord for the sample bottle is fully plugged into the back wall of the interior of
your unit.
2.
The sample bottle is securely mounted in the clip provided.
3.
The level of the propylene glycol in the bottle is full.
In the event that the bottle has broken during shipping, contact our service department for a
replacement bottle. Spare replacement bottles and clips are available as accessories.
Preprogrammed Settings
The monitoring system alarms are preset to common Federal Drug Administration (FDA) and
Center for Disease Control (CDC) industry standards for pharmaceutical drug and vaccine
storage conditions. Settings can be customized via the
Settings
and
Alerts
menus.
Temperature Display & Alarm
– The display is preset to read the glycol sensor but can
be changed to read the air sensor. Temperatures are detected and displayed in Celsius
units to one tenth of a degree. For example, 4.0 °C. Upper and lower alarm thresholds
are driven by the sensor selected for the display.
o
Set Point Temperature
- The unit preset/default is set at 4.0 °C (39.2 °F)
o
Upper Threshold Alarm
- The unit preset/default is set at 8.0°C (46.4 °F)
o
Lower Threshold Alarm
- The unit preset/default is set at 2.0°C (35.6 °F)