5.5 Controller Settings
The controller settings allow you to set the relay type and relay status of the controller.
ENT
SETUP
HOLD
2
ENT
SETUP
HOLD
3
SETUP
HOLD
1
SETUP
HOLD
2
ENT
SETUP
HOLD
3
ENT
ENT
From measurement mode
press ENT key
to enter setup mode as
described in section
1
1
5.1. The LCD shows the first screen of setup mode
(
rAng
).
Press
Ÿ
or
ź
key
to
select controller settings screen (
Cntr
).
Press ENT key
to access controller settings (
Cntr
).
Select the controller type
: The lower display shows ‘
tyP’
. The upper
display shows the last configured controller type. (‘OFF’ or ‘L.Ct’).
Press
Ÿ
or
ź
key
to
select the desired controller type.
2
OFF
:
Controller Off
Selecting ‘OFF’, the conductivity controller stops functioning as
a controller by preventing relay functions. Once the controller
type is set to ‘OFF’, the setup mode does not show you the
Relay 1/Relay 2 settings.
L.Ct
:
Limit value control (Controller On)
Selecting ‘L.Ct’, the conductivity controller is set to function as
a controller. When the controller type is set to ‘L.Ct’, the Relay
1/Relay 2 settings are available in the setup mode for you to
configure relay operating parameters.
Press ENT key
to confirm your selection.