Instruction Manual
D
lpha-CON 560
32
Other
Power Input
Uni9 V DC Adaptor and 24V DC Supply
Dimensions (W x H x D)
96mm x 96mm x 66 mm
Weight (Estimated)
210g
Ambient Temp. operating range
0 to 40
o
C
Maximum Relative Humidity
80% up to 31
o
C decreasing linearly to 50% at 40
o
C
Calibration Default Settings (CAL dEF)
Conductivity
Calibration factor
Reset to 1.000
Factory Default Settings (FCt dEF)
Conductivity
Calibration factor
Reset to 1.000
Cell constant
1.00
Cell type
2-cell
Measuring range
0 to 2000
μ
S/cm
Line resistance
0.000
Temperature
Unit of measurement
o
C
Temperature compensation
ATC On
Temperature coefficient
2.1% (Both process & calibration temp. coefficients)
Temperature coefficient type
Linear
MTC (ATC Off)
25
o
C (Both process & calibration temperature)
Controller Output
Controller type
Limit control
Relay status
De-energized
Relay 1 (SP1)
Set point value
100
μ
S
Relay function
Low
Hysteresis 20
μ
S
On-delay time
0 seconds
Off-delay time
0 seconds
Relay 2 (SP2)
Set point value
1900
μ
S
Relay function
High
Hysteresis 20
μ
S
On-delay time
0 seconds
Off-delay time
0 seconds
Display
Backlight On
www.
GlobalScientificSupply
.com
GLOBAL SCIENTIFIC SUPPLY
The Laboratory Supply Company!