5.6 Measuring Range Settings
The conductivity controller has 6 different measuring ranges. You need to select a
range which is the most suitable for conductivity reading of your applications/samples.
The factory default is 2000
μ
S (0 to 2000
μ
S/cm).
The measuring range you select here will affect the way the controller behaves in the
measurement mode and calibration mode. In measurement mode, if the conductivity
reading of the controller is above the selected measuring range, the LCD shows “Or”
(over range) error. In calibration mode, the controller adjusts calibration tolerance
based on full scale of the selected measuring range. The calibration is carried out
only for the selected measuring range.
WARNING:
If you change the measuring range of the controller after calibration is
done, the controller will reset back to its factory default calibration.
From measurement mode
press
ENT key
to enter setup mode as
described in section
1
1
SETUP
HOLD
ENT
1
μ
S
HOLD
SETUP
2
μ
S
HOLD
SETUP
2
μ
S
HOLD
SETUP
2
μ
S
HOLD
SETUP
2
mS
HOLD
SETUP
2
mS
HOLD
SETUP
2
ENT
ENT
ENT
ENT
ENT
ENT
5.1. The LCD
shows the first screen of setup mode
(
rAng
).
Press ENT key
to access
measurement range setting (rAng).
The upper display shows the
currently selected measuring range.
The factory default is 2000
μ
S. The
lower display shows the currently
configured cell constant (k). (Refer
section
2
4.2 for setting cell constant)
Press
Ÿ
or
ź
key
to select the
desired measuring range.
Press
ENT key
to confirm your selection.
The controller reverts to
rAng
screen.
Press
Ÿ
or
ź
key
to access other setup screens or
press
Ÿ
and
ź
key
simultaneously
(escape) to return to measurement mode.
Measuring range
Resolution
0.000 to 2.000
μ
S/cm 0.001
μ
S/cm
0.00 to 20.00
μ
S/cm 0.01
μ
S/cm
0.0 to 200.0
μ
S/cm 0.1
μ
S/cm
0 to 2000
μ
S/cm 1
μ
S/cm
0.00 to 20.00 mS/cm
0.01 mS/cm
0.0 to 200.0 mS/cm
0.1 mS/cm