5.4 Relay 1/Relay 2 Settings
If controller type is set to ‘OFF’ (Refer section 5.5 for controller settings),
the ‘Relay 1/Relay 2 Setting’ is not available.
NOTE:
The SP1 settings determine the operating parameters for Relay 1; while SP2 settings
define the operating parameters for Relay 2. Since these two settings have identical
setup parameters, only SP1 is described here.
Note that when you change measuring range of the controller, the values you have
set for SP1 & SP2 will not be affected. However, the unit of measurement &
resolution displayed in LCD for SP1 & SP2 will change according to the newly
selected measuring range.
SETUP
HOLD
1
ENT
SETUP
HOLD
2
ENT
SETUP
HOLD
SETUP
HOLD
3
3
ENT
ENT
SETUP
HOLD
4
ENT
SETUP
HOLD
SETUP
HOLD
ENT
ENT
5
6
μ
S
μ
S
μ
S
μ
S
From measurement mode
press ENT key
to enter setup mode as
described in section
1
1
5.1. The LCD shows the first screen of setup mode
(
rAng
).
Press
Ÿ
or
ź
key
to
select relay 1 (Set point 1) settings screen
(
SP1
).
Press ENT key
to access relay 1 settings (
SP1
).
Setting set point value
: The upper display shows the last configured set
point value, otherwise the default value 100
μ
S.
Press
Ÿ
or
ź
key
to
enter the value for set point 1, at which control function of the
conductivity controller should activate.
Press ENT key
to confirm your
set point value.
2
Select relay function
: The upper display shows the last configured relay
function (
LO
= Low or
HI
=High).
Press
Ÿ
or
ź
key
to select the desired
function.
3
Press ENT key
to confirm your selection.