Instruction Manual MM850404 ed.05
18
4 RECOMMENDATIONS
The analyzer should be used in environments where the temperature is between -5°C and +45°C.
When the analysis is complete, before turning the instrument off, remove the probe from the gas pathway and wait
about 30 seconds, cleaning the pneumatic circuit from gas.
Do not use the analyzer with clogged filters or filled with humidity.
Before placing the probe in the case, be sure that the probe, the water trap and the hose are clean and cold.
If you have ordered a double hose sampling probe, insert both connectors to the instrument during the gas analysis
(pay attention to color correspondecy).
To empty the water trap from the condensate water, remove the draining plug. DO NOT OPEN the water trap. DO
NOT leave the water trap without the draining rubber plug (cat. EE650081).
For the best efficiency and accuracy, we suggest to calibrate the instrument every year
.
WARNING
I
F THE INSTRUMENT IS STORED AT TEMPERATURE EXCEEDING THE OPERATIVE LIMITS
,
THE ANALYZER WILL A FEW NEEDS SOME
MINUTES TO WARM
-
UP TO THE ACTUAL AMBIENT TEMPERATURE
,
BEFORE STARTING THE OPERATION
.
4.1 Power
Supply
The 2000 series
can be powered from:
internal rechargeable Ni-MH battery.
External battery charger, supplied as a standard accessory (the batteries must be installed).
Optional external DC auto battery charger (the batteries must be installed).
The Ni-MH rechargeable batteries allows long operation and does not need maintenance. The same batteries power
both the instrument and the internal printer. The internal battery will grant 6 hours of continuous operation (without
printing and display back light off).
N
OTE
:
90%
BATTERY CHARGING CAN BE OBTAINED IN
8
HOURS WITH THE INSTRUMENT TURNED OFF
.
During operation a fully battery symbol “
≠
” will be displayed on the display. This symbol means that the batteries are
completely full. When the batteries will be discharged the symbol “
–
” will appear and the instrument still has about 20
minutes operation capability to end the running analysis.
The battery symbol indicates that a full charge is required. Use only the dedicated battery charger supplied by Eurotron
with the instrument.
CAUTION:
O
LD BATTERIES CAN LEAK AND CAUSE CORROSION
.
N
EVER LEAVE RUN DOWN BATTERIES IN THE INSTRUMENT
WARNING
T
HE INSTRUMENT IS SHIPPED WITH AN AVERAGE LEVEL OF BATTERY CHARGE
.
A
FTER UNPACKING
,
A FULL CHARGE OF THE BATTERY
IS RECOMMENDED
,
BY CONNECTING THE INSTRUMENT TO THE MAIN LINE THROUGH THE BATTERY CHARGER
(OFF
CONDITION
)
FOR
8-
10
HOURS AT A TEMPERATURE BETWEEN
10
AND
30
°C