Instruction Manual MM850404 ed.05
10
Accuracies and Ranges (EcoLine 2000 & GreenLine 2000)
Parametro
Sensor type
Range
Resol.
Max responce
Accuracy
O
2
electrochemical
from 0 to 25.0%
0.1%
20 sec.
±
0.2% Vol.
CO
H
2
compensated up
to 2000ppm
electrochemical
from 0 to 8000 ppm
1 ppm
50 sec.
±
10 ppm up to 300 ppm
±
4% rdg. up to 2000 ppm
±
10% rdg. >2000 ppm
NO
electrochemical
from 0 to 4000 ppm
1 ppm
50 sec.
±
5 ppm up to 125 ppm
±
4% rdg. up to 4000 ppm
CO
2
Calculated
from 0 to 99.9 %
0.1 %
Tair
Pt100
from -10 to 99.9 °C
0.1 °C
±
0.5 °C
Tgas
Tc K
from -10 to 600.0 °C
0.1 °C
±
1 °C
Δ
T
Calculated
Pressure / draft
Piezo
from -10 to +100hPa
0.01 hPa
±3Pa up to 300hPa
±
1% rdg elsewhere
Excess air
Calculated
from 1.00 to infinite
0.01
Efficiency
Calculated
from 0 to 99.9%
0.1 %
Smoke index
External pump
from 0 to 9
Technical units and ranges can be converted directly from ppm to mg/Nm
3
, mg/kwh and from hPa to mmH
2
O, mbar or
inH
2
O.
The relative accuracy shown are expressed as absolute or % of rdg errors at -5°C to +40°C ambient temperature.
The maximum response time shown is referred to 90% signal changes.
The pressure relative accuracy shown is valid only after the auto-zero procedure.
Specifications may change without notice.
Accuracies and Ranges (UniGas 2000)
Parametro
Sensor type
Range
Resol.
Max responce
Accuracy
O
2
electrochemical
from 0 to 25.0%
0.1%
20 sec.
±
0.2% Vol.
CO
H
2
compensated up
to 2000ppm
electrochemical
from 0 to 8000 ppm
1 ppm
50 sec.
±
20 ppm up to 300 ppm
±
4% rdg. up to 2000 ppm
±
10% rdg. >2000 ppm
CO
electrochemical
from 0 to 2000 ppm
1 ppm
50 sec.
±
20 ppm up to 400 ppm
±
5% rdg. up to 2000 ppm
CO
2
Calculated
from 0 to 99.9 %
0.1 %
Tair
Pt100
from -10 to 99.9 °C
0.1 °C
±
0.5 °C
Tgas
Tc K
from -10 to 600.0 °C
0.1 °C
±
1 °C
Δ
T
Calculated
Pressure / draft
Piezo
from -10 to +100hPa
0.01 hPa
±3 Pa up to 300 Pa
±
1% rdg elsewhere
Excess air
Calculated
from 1.00 to infinite
0.01
Efficiency
Calculated
from 0 to 99.9%
0.1 %
Smoke index
External pump
from 0 to 9
Technical units and ranges can be converted directly from ppm to mg/Nm
3
, mg/kwh and from hPa to mmH
2
O, mbar or
inH
2
O.
The relative accuracy shown are expressed as absolute or % of rdg errors at -5°C to +40°C ambient temperature.
The maximum response time shown is referred to 90% signal changes.
The pressure relative accuracy shown is valid only after the auto-zero procedure.
Specifications may change without notice.
Fuel Technical Data
The instrument includes as standard the technical data for 4 of the most common fuels.
Using the optional GasConfig software, it is possible to modify or add data of up to 10 different fuels.
1.3 Calibration
Certificate
Each portable gas analyzer
2000 series
, is factory calibrated and certified against
Eurotron
Standards, that are
periodically certified by an International recognized Laboratory, and shipped with a Report of Calibration stating the
nominal and actual values and the deviation errors.