error conditions. (e.g. O
2
>18%).
Excluded
CO sensor excluded by electro-valve
Automatic CO sensor protection is active.
Perform one or more autozero cycle in fresh
air.
Tc Error
The thermocouple is unconnected or
damaged.
Connect or replace the temperature probe.
Tsens Er
Cell compensation error
Return the instrument to Eurotron for
maintenance.
Zero Err
The autozero procedure is failed
Perform one or more autozero cycle in fresh
air. If the problem remains, Return the
instrument to Eurotron for maintenance.
Sens. Er
Broken cell
Return the instrument to Eurotron for
maintenance.
No meas.
Empty Memory
No sens.
The cell is non installed
No probe
No external probe disconnected on
mini-din connector.
Connect or replace the temperature probe.