Instruction Manual MM850404 ed.05
9
1.2 Specifications
Type
: 1 or 2 cells hand-held flue gas analyzer.
Calibration
: automatic calibration procedure at instrument switching-on.
Self-diagnosis
: Sensors efficiency test with status announcement.
Fuel types
: Up to 10 totally programmable.
Pump
: rate of flow 1.4 lit /min. @ 100mbar
Power supply
: High capacity Ni-MH rechargeable battery pack / external battery charger.
Charging time :
8h at 90% with instrument off.
Battery life
: 6h continuous operation (without printer and back-light).
Built-in printer
: optional impact type 24 columns with 58 mm large and 8 meters long paper roll
Built-in Printer power supply
: using the analyser battery pack.
Built-in Print autonomy
: up to 40 reports with battery full.
Internal data memory
: up to 250 complete analysis tests structured by Tags (max. 40).
Service and user data
: 3 programmable rows x 16 char for each programmed customer using a PC and GasConfig
Software.
Report Herder
: 4 rows x 16 char programmable from GasConfig Software
Display
: Large (40x56 mm) LCD display with automatic back-light device.
Serial interface
: Infrared interface for wireless HP82240B thermal printer. External RS232 adapter cable available
for PC communication.
Smoke measurement
: Using the optional external manual pump. Index memory store and printout capability as
standard.
Flue gas Probe
: stainless steel with rubber handle and incorporated temperature sensor. 2 meters long rubber hose
included.
Water trap
: external with purge plug.
Line filter
: with replaceable cartridge.
Optional probes
: ambient CO, explosive gas leakage sniffer.
Working temperature
: from –5°C to +45 °C (up to 50°C for short time)
Storage temperature
: from –20 to +60°C (3 months maximum at temperatures exceeding the operational limits).
Dimensions
: 2000: 105x75x290 mm; 2000+ 115x90x330mm
Weight
: 2000: 0.9 kg; 2000+ 1.1Kg, with battery and printer
External optional ambient CO probe
Range:
from 0 to 500ppm
Accuracy:
±
5ppm up to 100ppm;
±
5% up to 500ppm
Resolution:
1ppm
Response time:
30s (t90)
Waiting starting time:
30s
Working temperature
: from –5°C to +45 °C
External optional natural gas leak detection probe
Pre-heating time:
30s minimum
Alarms indication:
visual with 5 steps
5 alarms levels:
100, 200, 300, 400, 500 ppm
Acoustic Alarm indication:
respectively 1, 2, 3, 5 bip/s, continuous
Response time:
5s (t90)
Alarm levels Accuracy:
±
10% at 90 days