HYDRABEAUTY
25
6.4
Confirmation before reporting a fault
6. Maintenance
Problems
Cause of speculation
Check list
Power
Faulty power cable
connection
▪
Check
ON/OFF status
▪
Check connection of power cable
Fuse
▪
Replacement with new fuse
Vacuum
Faulty container
connection
▪
Check the connection of waste container
Faulty handpiece
connection
▪
Check the connection of handpiece with main body
Damage on handpiece
tip
▪
Check the damage of tip
Unusual
solution
Sparying
quantity
Inner system clogging
▪
Check the spraying quantity of solution1~3
.
If the spraying quantity of solution 1~3 is not appropriate, but
cleansing mode is usual, then should doubt the problem on
clogged inner system.
Clogged handpiece
▪
Check the spraying quantity of solution and vacuum
status by set both to maximum level and check for 20-
30sec
▪
Check the status of the device by processing cleansing
mode. If the cleansing mode is not working properly, then
should doubt clogged handpiece
Unusual
device
performance
Full waste container
▪
Check whether the container is full
When the container has filled more than 2.4L, there will be
warning message shown on the screen, and automatically the
device will be stopped.
Faulty power status
▪
Check
ON/OFF status
▪
Check connection of power cable