6.1 Outlines
6. Maintenance
1. Describes self-diagnostic method, including cleaning and storage method, and error code information to
maintain the device's performance.
6.2 Cleaning
1. Aqua handpiece valve management
① After completing the treatment, wash the machine in the automatic cleaning mode, C. Basic washing is
required to keep cleanliness from infection and germs.
2. Aqua handpiece management
① Aqua handpiece is disposable.
② Should waste the tip when it is used.
3. Large waste container
① After the treatment, use a cleaning tool to clean the waste solution from the container under running
water.
② when the large waste container is filled or more than once a day, the container should be cleaned
4. Diamond peeling tip & filter
① Remove the filter by separating the tip from handpiece and replace it with new one.
② The removed tip should be cleaned thoroughly with an ultrasonic cleaner or a soft brush.
5. Handpiece
① Wipe clean with wet towel and soft tissue after the treatment.
6. Main body
① Wipe the main body and screen clean with soft towel,
6.3 Storage
1. If the product should be stored for long periods of time, turn off the power and avoid humid places..
2. When storing the product, keep handpiece, tip, and solutions separately from the main body.
CAUTION
When maintaining the device, the device must be turned off and the power cable should be
disconnected.
Cleaning and storing with the power on, will cause serious damage or danger to the device.
CAUTION
The aqua hand piece tip is disposable and must be discarded after each treatment, and
no liability for any problems due to reuse is provided by the Eunsung Global Corp.