7.6.1. CALIBRATE pH 7/REF.POINT
The menu CALIBRATION allows the user to calibrate the instrument through the use of buffer solutions.
After press
CALIBRATE pH 7/REF. POINT
dipping the probe in the pH 7 buffer solution is the calibration
procedure’s first step. When the reading is stable, adjust the value by using the + and – keys and press
►
key to
confirm.
The next step is to put the probe into pH 4 or pH 9 buffer solution and press OK on
CALIBRATE BUFFER 4/9.
When the reading is stable, adjust the value by using the + and – keys and press
►
key to confirm.
The instrument is so calibrated.
For Redox and Chlorine, calibrate first calibration point by using buffer solution (e.g. 650mV) o by using a
photometer ( CALIBRATE BUFFER for REDOX and CALIBRATE DPD1 for CLORINE ).
Second calibration point (available in EXPERT mode only) is the "ZERO", which is possible to calibrate
by disconnect the probe and shortcutting BNC input connector.
7.7. MENU ALARM
Starting from the MAIN SCREEN press
OK
key than press
+
key to move the selector on
MENU ALARM
.
Press
OK
key to enter.
Press
OK
key to enter
into
MENU
MAIN SCREEN
620mV
SET1:OFF mA1:10.4
SET1:ON mA2:10.4
7.47 pH
SET1:ON mA1:12.4
SET1:OFF mA2:12.4
11.63
Cl ppm
SET1:OFF mA1:10.4
SET1:ON mA2:10.4
Summary of Contents for eSelect B2
Page 1: ...eSelect B2 B3 ...
Page 2: ......
Page 3: ...ESPANOL 3 ...
Page 7: ...3 3 CONNECTIONS 3 1 eSelect B2 POWER SUPPLY CARD CONNECTIONS ...
Page 8: ...4 3 2 eSelect B2 TTL OUTPUT CONNECTIONS ...
Page 9: ...5 3 3 eSelect B3 POWER SUPPLY CARD CONNECTIONS ...
Page 10: ...6 3 4 eSelect B3 TTL OUTPUT CONNECTIONS ...
Page 12: ...8 4 CONTROL PANEL DESCRIPTION 4 1 eSelect B2 CONTROL PANEL 4 2 eSelect B3 CONTROL PANEL ...
Page 14: ...10 5 DIMENSIONS ...
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ...COD DMU00193ML1 A 09 2013 ...