19
MENU SETUP
1/2
MENU MODE
EXPERT
METER TYPE
pH
T SENSOR
NONE
SET TEMPERATURE
25.0
CALIBRATION MENU TIMEOUT
05:00
DELAY AT STARTUP
00:05
AUX OUTPUT MODE
OFF
YEAR 2013
MONTH 5
DAY 15
7.4.7. AUX OUTPUT MODE
The controller has a calendar and an internal clock for the management of the exits with timer and the storage of
the data recorded by the controller; in order to set up the clock the following settings must be carried out.
Once the date and time are set, by pressing the
+
or
-
key, the choice are the follow:
OFF
function: the aux output is deactivated;
Cleaning
function: it is possible to activate a dosing pump for the cleaning of the electrode, up to a
maximum of 4 interventions (timer and scheduled mode) during the day; the difference from the
flocculant mode, is that the Cleaning mode interrupts the tool’s operations (disabling the set-point). At
the end of such intervention the tool awaits the start up time (see DELAY AT STARTUP).
Flocculant
function: it is possible to activate the flocculant dosing system, up to a maximum of 4
interventions (timer and scheduled mode) during the day.
Starting from the
MENU SETUP
screen, enter in
AUX OUTPUT MODE
by pressing
+
key or
–
key.
After selected
AUX OUTPUT MODE
press
OK
key to set between
OFF
,
CLEANING
and
FLOCCULANT
.
Once one of the two functions is activated (
flocculant
or
cleaning
) it is necessary to set the auxiliary outputs
that need to be activated by select the days in which the auxiliary output needs to be activated.
7.4.7.1. CLEANING AUX OUTPUT MODE
Starting from the
MENU SETUP
screen, and selected CLEANING as
AUX OUTPUT MODE
press
OK
key.
Than Set
CYCLE TIME
,
WEEKDAYS1
,
WEEKDAYS2
,
WEEKDAYS3
,
WEEKDAYS4
as follow:
CYCLE TIME
: define how long time the AUX OUTPUT relay have to be activated;
WEEKDAYS1
,
WEEKDAYS2
,
WEEKDAYS3
,
WEEKDAYS4
: define the days of the week in wich the
aux output have to be activated; after setting the
WEEKDAYS
on the display appear to set the
START
TIME
7.4.7.2. FLOCCULANT AUX OUTPUT MODE
Starting from the
MENU SETUP
screen, and selected FLOCCULANT as
AUX OUTPUT MODE
press
OK
key.
Than Set
CYCLE TIME
,
WEEKDAYS1
,
WEEKDAYS2
,
WEEKDAYS3
,
WEEKDAYS4
as follow:
CYCLE TIME
: define how long time the AUX OUTPUT relay have to be activated;
WEEKDAYS1
,
WEEKDAYS2
,
WEEKDAYS3
,
WEEKDAYS4
: define the days of the week in wich the
aux output have to be activated; after setting the
WEEKDAYS
on the display appear to set the
START
TIME
Press
+
or
–
to choose between
OFF
,
CLEANING
and
FLOCCULANT
.
Summary of Contents for eSelect B2
Page 1: ...eSelect B2 B3 ...
Page 2: ......
Page 3: ...ESPANOL 3 ...
Page 7: ...3 3 CONNECTIONS 3 1 eSelect B2 POWER SUPPLY CARD CONNECTIONS ...
Page 8: ...4 3 2 eSelect B2 TTL OUTPUT CONNECTIONS ...
Page 9: ...5 3 3 eSelect B3 POWER SUPPLY CARD CONNECTIONS ...
Page 10: ...6 3 4 eSelect B3 TTL OUTPUT CONNECTIONS ...
Page 12: ...8 4 CONTROL PANEL DESCRIPTION 4 1 eSelect B2 CONTROL PANEL 4 2 eSelect B3 CONTROL PANEL ...
Page 14: ...10 5 DIMENSIONS ...
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ...COD DMU00193ML1 A 09 2013 ...