18
MENU SETUP
1/2
MENU MODE
EXPERT
METER TYPE
pH
T SENSOR
NONE
SET TEMPERATURE
25.0
CALIBRATION MENU TIMEOUT
05:00
DELAY AT STARTUP
00:05
AUX OUTPUT MODE
OFF
YEAR 2013
MONTH 5
DAY 15
MENU SETUP
1/2
MENU MODE
EXPERT
METER TYPE
pH
T SENSOR
NONE
SET TEMPERATURE
25.0
CALIBRATION MENU TIMEOUT
05:00
DELAY AT STARTUP
00:05
AUX OUTPUT MODE
OFF
YEAR 2013
MONTH 5
DAY 15
MENU SETUP
1/2
MENU MODE
EXPERT
METER TYPE
pH
T SENSOR
NONE
SET TEMPERATURE
25.0
CALIBRATION MENU TIMEOUT
05:00
DELAY AT STARTUP
05:00
AUX OUTPUT MODE
OFF
YEAR 2013
MONTH 5
DAY 15
If PT100 will be set a PT100 sensor have to be connected to the dedicated connector. Selecting PT function, the
temperature measured by PT100 probe is shown on the display (while the visualization of the measure is in
progress).
If
NONE
will be set it is possible to set the real temperature as show in next paragraph.
7.4.4. SET TEMPERATURE
Starting from the
MENU SETUP
screen
,
if the PT100 sensor is not setted, enter in
SET TEMPERATURE
menu
by pressing
+
key or
–
key.
After selected
SET TEMPERATURE
press
OK
key to set the temperature value.
7.4.5. CALIBRATION MENU TIMEOUT
It represent the exit delay time from the programming menu in case no keys are pressed during the probe
calibration stage.
Starting from the
MENU SETUP
screen, enter in
CALIBRATION MENU TIMEOUT
by pressing
+
key or
–
key.
After selected
CALIBRATION MENU TIMEOUT
press
OK
key to set the time value.
7.4.6. DELAY AT STARTUP
It define the delay time of measurement actuation from switching on the instrument.
Starting from the
MENU SETUP
screen, enter in
DELAY AT STARTUP
by pressing
+
key or
–
key.
After selected
DELAY AT STARTUP
press
OK
key to set the time value.
Press
+
or
–
to choose the
temperature value.
Press
+
or
–
to choose the time
value (minutes and seconds).
Press
+
or
–
to choose the time
value (minutes and seconds).
Summary of Contents for eSelect B2
Page 1: ...eSelect B2 B3 ...
Page 2: ......
Page 3: ...ESPANOL 3 ...
Page 7: ...3 3 CONNECTIONS 3 1 eSelect B2 POWER SUPPLY CARD CONNECTIONS ...
Page 8: ...4 3 2 eSelect B2 TTL OUTPUT CONNECTIONS ...
Page 9: ...5 3 3 eSelect B3 POWER SUPPLY CARD CONNECTIONS ...
Page 10: ...6 3 4 eSelect B3 TTL OUTPUT CONNECTIONS ...
Page 12: ...8 4 CONTROL PANEL DESCRIPTION 4 1 eSelect B2 CONTROL PANEL 4 2 eSelect B3 CONTROL PANEL ...
Page 14: ...10 5 DIMENSIONS ...
Page 32: ......
Page 33: ......
Page 34: ......
Page 35: ...COD DMU00193ML1 A 09 2013 ...