2
It is also possible to place it directly in the axis search position by clicking once on the value of
the cylinder axis for the eye concerned.
Ask the patient the following question:
"
Look at the dots. Tell me if they look sharper, darker, more contrasted in position 1, position 2 or if they
look identical to you?
“
To:
Show the dots in position 1, press the "1" key
on the console keyboard.
To show the dots in position 2, press, press the
"2" key on the console keyboard.
It is important to always propose the three options:
Position 1
Position 2
Same
Blue highlighting of positions 1 and 2
Cross cylinder position change
The cross cylinder is placed in the cylinder axis verification position, oriented according to the direction
of the negative axis of the patient’s correction cylinder.
This axis is visually represented by the black line below.
The white dots represent the positive axis.
>
The position change appears in the test presentation area in two ways:
>
U
SER
MANUAL
> P
ERFORMANCE
OF
TESTS
DURING
A
REFRACTION
EXAMINATION
Vision-R 700 - Essilor automatic phoropter > V1 - 09-2020
56
Summary of Contents for VISION-R 700
Page 1: ...USER MANUAL...
Page 4: ...I INTRODUCTION...
Page 8: ...II INSTRUMENT...
Page 16: ...III ADJUSTMENTS BEFORE THE EXAMINATION...
Page 24: ...IV BASIC FUNCTIONS FOR PERFORMING A REFRACTION EXAMINATION...
Page 37: ...V PERFORMANCE OF TESTS DURING A REFRACTION EXAMINATION...
Page 76: ...VI MEASUREMENT...
Page 80: ...VII REFRACTION PROGRAMS...
Page 93: ...VIII INSTRUMENT SETTINGS...
Page 109: ...IX TROUBLESHOOTING...
Page 111: ...X TECHNICAL DATA...
Page 114: ...XI CAUTIONS WARNINGS...
Page 124: ...XII QR CODE...
Page 125: ...USER MANUAL QR CODE 125 Vision R 700 Essilor automatic phoropter V1 09 2020...
Page 126: ...USER MANUAL QR CODE Vision R 700 Essilor automatic phoropter V1 09 2020 126...
Page 127: ...USER MANUAL QR CODE 127 Vision R 700 Essilor automatic phoropter V1 09 2020...