Electromagnetic Immunity for All Medical Equipment
The
MyLab
system is intended for use in the electromagnetic environment specified below. The
customer or the user of the
MyLab
system should assure that it is used in such an environment.
Immunity Test
IEC60601
Test Level
Compliance
Level
Electromagnetic Environment and
Measures to Be Taken
Electrostatic
discharge (ESD)
IEC 61000-4-2
6 kV on
contact
8 kV in air
6 kV on
contact
8 kV in air
The floor should be in wood, concrete
or ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least at 30%.
Electrical fast
transient/burst
IEC 61000-4-4
2 kV for
power supply
lines
1 kV for
input/output
lines
2 kV for
power supply
lines
1 kV for
input/output
lines
Mains power quality should be that of a
typical commercial or hospital
environment.
Surge
IEC 61000-4-5
1 kV
differential
mode
2 kV
common
mode
1 kV
differential
mode
2 kV
common
mode
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% of rated
voltage (
U
T
)
(voltage gap
>95%)
for 0,5 cycle
40%
U
T
(voltage dip
60%)
for 5 cycles
70%
U
T
(voltage dip
30%)
for 25 cycles
<5%
U
T
(voltage dip
>95%)
for 5 sec
<5% of rated
voltage (
U
T
)
(voltage gap
>95%)
for 0,5 cycle
40%
U
T
(voltage dip
60%)
for 5 cycles
70%
U
T
(voltage dip
30%)
for 25 cycles
<5%
U
T
(voltage dip
>95%)
for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the
MyLab
system requires continued operation
during power mains interruptions, it is
recommended that the
Mylab
system be
powered from an uninterruptible power
supply or a battery.
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level
Summary of Contents for 350009500
Page 1: ...Rev 02 April 2016 MyLabDelta GETTING STARTED 350009500 ...
Page 10: ...MyLab G E T T I N G S T A R T E D 6 ...
Page 17: ...MyLab G E T T I N G S T A R T E D 1 7 GETTING STARTED ...
Page 18: ...MyLab G E T T I N G S T A R T E D 1 8 ...
Page 19: ...MyLab G E T T I N G S T A R T E D 1 9 GETTING STARTED ...
Page 26: ...MyLab G E T T I N G S T A R T E D 1 16 ...
Page 34: ...MyLab G E T T I N G S T A R T E D 2 8 ...
Page 40: ...MyLab G E T T I N G S T A R T E D 3 6 ...
Page 68: ...MyLab G E T T I N G S T A R T E D 5 12 ...
Page 76: ...MyLab G E T T I N G S T A R T E D 6 8 ...
Page 88: ...MyLab G E T T I N G S T A R T E D 7 12 ...