User Manual
Start up
Calibrating the electronic anemometer
1- Press silmutaneously the main switch while switching on the main swith on the back of the control panel (off:O, ON:I)
2- Release immediately the mute button. A beep will proove the unit has started the calibration mode. LEDS lights will blink
fastly without any sound during 5min. 30 sec. LEDs lights blinking is different when an alarm is triggered.
3- Light blinking stops. The anemometer is calibrated and the unit is operational.
4- Turn on fan of the unit on the control panel.
5- Green light and LEDs lights are lighting.
Good practices :
- If you are using prefilters, please be sure they are correctly located
- If you are using Reverso sash (Captair Smart 392 or 483 or 633 or 714 ductless fume hoods), please do the calibration as
indicated.