21
User Manual
Procedure for automatic detection for the saturation of a
molecular filter using acetone
Acetone evaporates easily and quickly at room temperature and can be used as a tracer to check mocular filters saturation.
Following procedure provides three methods to determine activated carbon filters saturation, evaporating a small amount of
acetone withing the Captair Smart fume hood and using either a colorimetric tube for acetone, Molecode S saturation sensor
or a PID device. The result of the test will indicate weither it is necessary to replace filter or not.
Required material:
•
5 mL of acetone (not provided)
•
Absorbing cloth (not provided)
•
Cup or crucible (not provided)
•
Hose (provided)
•
Sampling pump (not provided)
•
Colorimetric tube (not provided)
Remove protection covers
Turn on the unit and start the air flow
Not provided materials can be sourced from various manufacturers, such as Gastec, Draeger,
RAE System or authorised distributors.
Always use gas sampling pump together with detector tube from the same brand
Ref. Gastec tube: 151 L (pump 2 times)
Ref. Draeger tube: 81 03 381 (pump 1 time)
Ref. RAE tube: 10-111-40 (sample 100cc)
Saturation detection using colorimetric tubes