24
Captair Smart
Pour 5 mL of acetone on an absorbing cloth placed
within the fume hood
Connect to the appropriate fan box depending on your unit model:
Carbon filter
fan box
Hose connection
Remove sample port cork and connect hose
to the fan box
321
392
483 / 633
714
Connect hose to the PID sampling probe
Select ‘acetone’ to be detected in PID menu (see user manual
if necessary).
If acetone concentration is more than to 5 ppm, change filter(s)
-European
fan connection:
- North American
fan connection: