23
User Manual
Proceed with air sampling following manufacturer’s recommendation
If test result is positive to acetone, change filter(s)
Saturation detection using PID Device:
Required material :
•
5 mL of acetone (not provided)
•
Absorbing cloth (not provided)
•
Cup or crucible (not provided)
•
Hose (provided)
•
PID device (not provided)
Materials not provided can be sourced from various RAE System manufacturer or authorised distributors.
Remove protection covers
Turn on the unit and start air flow