37
User Manual
Replacement frequency of filtration breakthrough sensor,
Molecode option
Molecular filters saturation detection alarm – Molecode (S : solvents or A : acids or F: formaldehyde - fitted as an integrated
option in your unit, work using semi-conductor or electrochemical sensor that must be changed every :
5 years for the Molecode S
2 years for Molecode A and F
After this period, Erlab cannot guarantee performances stability nor detection sensitivity.
Equipment replacement requires the intervention of a qualified engineer, able to replace internal components and perform
system new configuration. The change requires less than one hour.
Erlab Maintenance department and accredited partners can provide this service. (service not available in all countries)
For further information, please get in touch with your contact person or Erlab dealer/distributor or visit our website
Admissible weighs on working surfaces
Tempered glass
work top
Stainless steel
work top
Trespa
®
Top Lab
PLUS
work top
50 kg/m
2
1187 lb/ft
2
110 kg/m
2
2610 lb/ft
2
110 kg/m
2
2610 lb/ft
2
Captair Smart 321
15 kg
33 lbs
35 kg
77 lbs
35 kg
77 lbs
Captair Smart 391
20 kg
44 lbs
45 kg
99 lbs
50 kg
110 lbs
Captair Smart 481
25 kg
55 lbs
65 kg
143 lbs
65 kg
143 lbs
Captair Smart 392
20 kg
44 lbs
60 kg
132 lbs
60 kg
132 lbs
Captair Smart 483
30 kg
66 lbs
80 kg
176 lbs
80 kg
176 lbs
Captair Smart 633
40 kg
88 lbs
100 kg
220 lbs
100 kg
220 lbs
Captair Smart 714
45 kg
99 lbs
115 kg
254 lbs
115 kg
254 lbs