6. Important hints

E

07

*

The cuff pressure must not exceed 300 mmHg

*

The measuring time should not exceed 2 minutes

*

A repeated measure is to be taken only after 2 min. rest

*

Protect the instrument from breaking and heavy vibration

*

Protect the instrument from dirt and moisture

*

Avoid damaging the rubber parts from sharp or pointed objects

*

Store the instrument at a temperature between 10 and 40°C and a relative air
humidity below 85 %

*

Do not expose the instrument, in particular the rubber parts, to direct ultraviolet
radiation

*

The cuff is to be used with a hygienic protective strip or it must be cleaned
before it is used for the next patient.

7. Maintenance and care

a) Manometer

In principle the manometer is maintenance-free. Dust particles in the air can block
the filter in the main body after a certain time. The bulb can be removed for cleaning
or replacement.

b) Cuff

Before cleaning the cuff, the bag must be removed. Covers with velcro fastening are
to be folded before washing so that the parts of fleece and of hooks are well
attached.

Hints for care

Non-shrink popelin

100% cotton

colour: green

black

Washing

60° mild washing

60° standard

Ironing position

weak

normal

Dry cleaning

as delicate clothing

as norm. clothing

Tumble-Dryer

mild

not possible

easy to wipe PUR-laminated material
all colours

Wet wiping

with a damp cloth,if necessary using a
mild washing liquid

Dry cleaning

as delicate clothing

Disinfecting

The cuffs have been tested to withstand the following
recommended disinfectants:
Cidex, Sporiciden, Microzid, Isopropyl-
alcohol 70%, Ethanol 70%, Buraton liquid.
After washing in disinfectants the cuff cover
should be rinsed and allowed to air dry.

Sterilization

The cuff has been tested to withstand
Ethylene Oxide (EtO) gas sterilisation. Be sure that all
safety precautions regarding
aeration after EtO exposure are followed.

!!! Never autoclave !!!

Nylon / all colours

Washing

30° mild w mild detergent

Wiping

with sponge or soft brush + mild
detergent, rinse with water

Summary of Contents for Aneroid

Page 1: ...truction manual Mo d ed e mp l o i Instrucciones de manejo I s t r u z i o n i p e r l u s o Aneroid Blutdruckmessger te Aneroid Sphygmomanometers Aneroid Tensiom tres Aneroid Esfigmoman metros Aneroi...

Page 2: ...ce apr s vente et r talonnage 12 Tabla de materias ES 1 Declaraci n de garant a 13 2 Descripci n del aparato 13 3 Preparaci n de la medici n 13 4 Realizaci n de la medici n 14 5 Elecci n del tama o de...

Page 3: ...sung von 2 mmHg Die Messgenauigkeit ber den gesamten Messbereich betr gt 3 mmHg Jedes Manometer wird kalibriert gegen unberechtigte Eingriffe gesichert und einer 100 igen Endkontrolle unterzogen 3 Vo...

Page 4: ...r mit automatischem Ventil Die Handhabung dieses Ger tes ist einfacher und komfortabler Durch Dr cken des Gebl seballs springt der sechseckige Knopf in der Mitte des Ventils heraus Der Messvorgang ist...

Page 5: ...nschette zuerst die Blase entfernen Bez ge mit Rapidverschluss vor dem Waschen falten so dass Flausch und Hakenband geschlossen aufeinander liegen Alle Metallteile sollen sich innerhalb der Manschette...

Page 6: ...en als fehlerhaft und m ssen sofort nachkalibriert werden Das Herstellungsjahr des Ger tes l sst sich aus den ersten beiden Ziffern der Seriennummer auf der Skala erkennen F r das Modell Switch ist ei...

Page 7: ...of 2 mmHg The measuring accuracy over the total measuring range is 3 mmHg After calibration and a 100 final control each manometer is sealed against unauthorised opening 3 Preparations for measuring...

Page 8: ...of inscribing the various color labels attached refer to item 8 b manometer with automatic valve The handling of this instrument is easier and more comfortable By pumping the bulb the hexagonal button...

Page 9: ...he cuff the bag must be removed Covers with velcro fastening are to be folded before washing so that the parts of fleece and of hooks are well attached Hints for care Non shrink popelin 100 cotton col...

Page 10: ...o be recalibrated The date of manufacture is shown by the first two numbers of the serial number A recalibration of the model SWITCH is not intended The faulty manometer housing can be replaced anywhe...

Page 11: ...n pr muni contre toute op ration non autoris e et soumis un contr le final de 100 3 Pr paration de la mesure Important Choisissez la taille correcte de la manchette Comme standard l appareil est fourn...

Page 12: ...ge ci jointes sont utilis es b pour tensiom tres valve automatique Le maniement de cet appareil est plus simple et confortable En pressant la poire le bouton hexagonal au centre de la soupape merge Le...

Page 13: ...e la partie frise et la partie aux petits crochets soient attach es Toutes les parties m talliques doivent tre l int rieur de la manchette Instructions Textile m tisse Coton 100 de lavage couleur vert...

Page 14: ...res du num ro de s rie sur le cadran P o u rl emo d l eS w i t c hu nr e c a l i b r a g en e s t p a sp r v u D a n sc ec a sl ab o t ed e mesure d fectueuse est directement remplac e sur place par u...

Page 15: ...g La exactitud de medici n en todo el alcance de medici n es de 3 mmHg Cada man metro es calibrado protegido contra intervenciones no autorizadas y sometido a un riguroso y completo control final 3 Pr...

Page 16: ...de flecha m nimamente hacia la izquierda m s lento m nimamente hacia la derecha m s r pido Tenga cuidado para que el tornillo de regulaci n no se gire con fuerza m s de lo que los topes derecho e izq...

Page 17: ...ro Lavar 60 prog ropa delicata 60 prog normal Planchado ajuste tibio normal limpieza en seco como prendas delicadas c prendas normal Secadora tambor con cuidado no es posible Tejido recubierto de PUR...

Page 18: ...inmediato El a o de fabricaci n del aparato corresponde a las dos primeras cifras del n mero de serie marcado sobre la esfera El modelo Switch no se puede calibrar Cajas medidoras con defecto pueden r...

Page 19: ...rotetto contro interventi non autorizzati e sottoposto ad un controllo finale completo al 100 3 Preparativi per la misurazione Importante La misura del bracciale deve essere esatta L a p p a r e c c h...

Page 20: ...c o n f o r t e v o l e P r e me n d ol ap e r a e s c e il pulsante esagonale che si trova nel centro della valvola Si esegue la misurazione come descritto sotto punto 4a Si regola una volta per tutt...

Page 21: ...ali con chiusura rapid sovrapponendo il nastro peloso soffice a q u e l l o c o n ig a n c e t t i T u t t e l e p a r t ii n me t a l l o d e v o n o t r o v a r s ia l l i n t e r n o d e l braccial...

Page 22: ...perfetti e devono essere subito ricalibrati Si pu r i l e v a r el a n n od i p r o d u z i o n ed a l l ep r i med u ec i f r ed e l n u me r od i serie sulla scala Per il modello Switch non prevista...

Page 23: ...20 220 180 180 170 122 Zubeh r Manschette Accessory Cuff Accessoire Manchette Accesorios manguito Accessori bracciale Doppelschlauch Double tube flexible double doble tubo due tubi uniti 1 Schlauch 1...

Page 24: ...loi et toute op ration nonautoris e Tarjeta de Garantia Concedamos sobre este aparato una garant a de 24 meses Excepci n Perfect Aneroid 36 meses a partis de la fecha de compra que cubre defector de f...

Reviews: