141
/ 266
3
CIRCUIT DESCRIPTION
Art. No.: 801
16-201
0
9
/ 200
4
an inductivity of the coils results which limits the high-frequency leakage current with this inductive
impedance.
Monitoring the high-frequency leakage currents
As intended, the high-frequency current should only be used for cutting or coagulating tissue. To do this, it
must flow to the active electrode in a small area and with a high current density, and flow back again from
the neutral electrode over a large area and with a low current density.
However, if the body of the patient is in contact with another electrically conductive object, the current may
also take an undesirable path and, for example, cause a burn there. These undesirable currents are known as
“leakage currents”. It is therefore important to know whether such a leakage current is flowing and how
large it is.
Such leakage current cannot simply disappear, but must instead always flow back to the generator. The
largest proportion of these high-frequency leakage currents flow back to the generator via the power cable
and the grounded conductor.
Since all power lines and the grounded conductor are directed through the ring core of the coil L1, a
corresponding high frequency voltage is induced if HF leakage current is present in the coil L1, which is
rectified with diode D1, restricted with D6, D7 and smoothed in capacitor C13. For further processing, this
DC voltage is fed to the processor system via the isolation amplifier IC12.
Changeover between flowing and capacitance-grounded output
High-frequency surgical units are generally available either in “capacitance ground of the neutral electrode”
or “floating neutral electrode” connection systems, whereby no components are integrated between the
connection to the neutral electrode and chassis in the latter connection system.
Both connection systems have advantages and disadvantages.
The advantage of capacitance ground is that the feed line from the neutral electrode to the patient is grounded
at a high frequency via the capacitor and the risk of burns is reduced. However, it is a disadvantage that
impermissible low-frequency leakage currents may flow through this capacitor during cardiac operations.
If the neutral electrode is operated as floating, on the other hand, the low-frequency leakage currents can be
kept very easily below the limit value of 10 or 50 mA during cardiac operations, but the risk of burns is
increased in this connection system since the line from the neutral electrode is about half the high frequency
voltage of the generator.
With the ICC high-frequency surgical units, the low-frequency leakage current to the patient is measured.
At the first malfunction, this must not exceed 50 mA. Once this condition is ensured, the unit is permitted
to carry the CF designation (Cardiac Floating). If the possible low-frequency leakage current is under 500
mA in the first case of malfunction, the unit is therefore not suitable for operations on the heart and has the
designation BF (Body Floating).
Initially the ICC is capacitance-grounded via capacitors C30 and C31 and the relay contact REL 2. The
advantages of capacitance ground are fully exploited. The low-frequency leakage current is measured via
the prescribed simulation R25, C28. If for any reason a low-frequency leakage current should appear, it is
then evaluated. If it exceeds the limit of 50 mA, the relay contacts REL 2 are then opened and thus a low-
frequency leakage current can no longer flow. The ICC thus fulfills the CF requirement for cardiac surgery
Motherboard
Summary of Contents for ICC 200
Page 1: ...Service manual ERBE ICC 200 ICC 300 H E ICC 350 09 2004...
Page 2: ......
Page 6: ...6 266...
Page 7: ...Chapter1 Test programs and adjustments...
Page 8: ......
Page 42: ...1 TEST PROGRAMS AND ADJUSTMENTS 42 266...
Page 52: ...1 TEST PROGRAMS AND ADJUSTMENTS 52 266...
Page 53: ...1 TEST PROGRAMS AND ADJUSTMENTS 53 266 Art No 80116 201 09 2004 Adjustment 1...
Page 56: ...1 TEST PROGRAMS AND ADJUSTMENTS 56 266...
Page 57: ...1 TEST PROGRAMS AND ADJUSTMENTS 57 266 Art No 80116 201 09 2004 Adjustment 2...
Page 60: ...1 TEST PROGRAMS AND ADJUSTMENTS 60 266...
Page 61: ...1 TEST PROGRAMS AND ADJUSTMENTS 61 266 Art No 80116 201 09 2004 Adjustment 3...
Page 64: ...1 TEST PROGRAMS AND ADJUSTMENTS 64 266...
Page 65: ...1 TEST PROGRAMS AND ADJUSTMENTS 65 266 Art No 80116 201 09 2004 Adjustment 4...
Page 68: ...1 TEST PROGRAMS AND ADJUSTMENTS 68 266...
Page 69: ...1 TEST PROGRAMS AND ADJUSTMENTS 69 266 Art No 80116 201 09 2004 Adjustment 5...
Page 72: ...1 TEST PROGRAMS AND ADJUSTMENTS 72 266...
Page 73: ...1 TEST PROGRAMS AND ADJUSTMENTS 73 266 Art No 80116 201 09 2004 Adjustment 6...
Page 76: ...1 TEST PROGRAMS AND ADJUSTMENTS 76 266...
Page 77: ...1 TEST PROGRAMS AND ADJUSTMENTS 77 266 Art No 80116 201 09 2004 Adjustment 7...
Page 79: ...1 TEST PROGRAMS AND ADJUSTMENTS 79 266 Art No 80116 201 09 2004 Adjustment 7 Phase angle cos j...
Page 80: ...1 TEST PROGRAMS AND ADJUSTMENTS 80 266...
Page 81: ...1 TEST PROGRAMS AND ADJUSTMENTS 81 266 Art No 80116 201 09 2004 Adjustment 8...
Page 83: ...1 TEST PROGRAMS AND ADJUSTMENTS 83 266 Art No 80116 201 09 2004 Adjustment 8 Spark monitor...
Page 84: ...1 TEST PROGRAMS AND ADJUSTMENTS 84 266...
Page 85: ...1 TEST PROGRAMS AND ADJUSTMENTS 85 266 Art No 80116 201 09 2004 Adjustment 9...
Page 88: ...1 TEST PROGRAMS AND ADJUSTMENTS 88 266...
Page 89: ...1 TEST PROGRAMS AND ADJUSTMENTS 89 266 Art No 80116 201 09 2004 Adjustment 10...
Page 92: ...1 TEST PROGRAMS AND ADJUSTMENTS 92 266...
Page 93: ...1 TEST PROGRAMS AND ADJUSTMENTS 93 266 Art No 80116 201 09 2004 Adjustment 11...
Page 96: ...1 TEST PROGRAMS AND ADJUSTMENTS 96 266...
Page 97: ...1 TEST PROGRAMS AND ADJUSTMENTS 97 266 Art No 80116 201 09 2004 Adjustment 12...
Page 100: ...1 TEST PROGRAMS AND ADJUSTMENTS 100 266...
Page 101: ...1 TEST PROGRAMS AND ADJUSTMENTS 101 266 Art No 80116 201 09 2004 Adjustment 13...
Page 104: ...1 TEST PROGRAMS AND ADJUSTMENTS 104 266...
Page 110: ...1 TEST PROGRAMS AND ADJUSTMENTS 110 266...
Page 114: ...1 TEST PROGRAMS AND ADJUSTMENTS 114 266...
Page 122: ...1 TEST PROGRAMS AND ADJUSTMENTS 122 266...
Page 127: ...Chapter2 ERROR list...
Page 128: ......
Page 136: ...2 ERROR LIST 136 266...
Page 137: ...Chapter3 Circuit description...
Page 138: ......
Page 168: ......
Page 169: ...Chapter4 Block diagrams...
Page 170: ......
Page 174: ......
Page 175: ...Chapter5 Circuit diagrams...
Page 176: ......
Page 177: ...PCBs for ICC 200 300 350...
Page 181: ...5 CIRCUIT DIAGRAMS 181 266 Art No 80116 201 09 2004 ICC 200 300 350 CPU PCB bare 40128 026...
Page 188: ...188 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Control board bare 40128 095...
Page 190: ...190 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 QC Power Stage bare 40128 127...
Page 192: ...192 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Power Module bare Power Module UL bare 40128 117...
Page 194: ...194 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 ST Power Stage bare 40128 129...
Page 198: ...198 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Senso board bare 40128 130...
Page 200: ...200 266 5 CIRCUIT DIAGRAMS ICC 200 300 350 Upper Wiring Module bare 40128 118...
Page 201: ...PCBs only for ICC 200...
Page 202: ...202 266 5 CIRCUIT DIAGRAMS...
Page 208: ...208 266 5 CIRCUIT DIAGRAMS ICC 200 Motherboard bare 40128 025...
Page 210: ...210 266 5 CIRCUIT DIAGRAMS ICC 200 Mono Output bare 40128 096...
Page 211: ...PCBs only for ICC 300...
Page 212: ...212 266 5 CIRCUIT DIAGRAMS...
Page 218: ...218 266 5 CIRCUIT DIAGRAMS ICC 300 Motherboard bare 40128 092...
Page 220: ...220 266 5 CIRCUIT DIAGRAMS ICC 300 Relay Board bare 40128 093...
Page 221: ...PCBs only for ICC 350...
Page 222: ...222 266 5 CIRCUIT DIAGRAMS...
Page 229: ...5 CIRCUIT DIAGRAMS 229 266 Art No 80116 201 09 2004 ICC 350 Motherboard bare 40128 092...
Page 231: ...5 CIRCUIT DIAGRAMS 231 266 Art No 80116 201 09 2004 ICC 350 Relay Board bare 40128 093...
Page 234: ...234 266 5 CIRCUIT DIAGRAMS ICC 350 Neurotest Neurotest Board 40128 027...
Page 236: ...236 266 5 CIRCUIT DIAGRAMS ICC 350 Neurotest Motherboard Neurotest bare 40128 028...
Page 238: ...238 266 5 CIRCUIT DIAGRAMS ICC 350 MIC MIEN DOKU MIC Board bare 40128 064...
Page 240: ...240 266 5 CIRCUIT DIAGRAMS ICC 350 MIC MIEN DOKU Protection Board bare 40128 141...
Page 241: ...AppendixA Part Numbers...
Page 242: ......
Page 263: ...Appendix B Abbreviations Notes Addresses...
Page 264: ...264 266 B APPENDIX B...
Page 266: ...Your notes...