21
•
Machine cleaning and disinfection in a G 7836 CD disinfector
(Miele & Cie. KG, Gütersloh, Germany) with the cleaning
agent neodisher ® mediclean forte (Dr. Weigert GmbH & Co.
KG, Hamburg, Germany) with the recommended washing
program.
5.4 Recommended methods
Erbe recommends the preparation procedures described below.
Equivalent different procedures are possible if not explicitly ex-
cluded. It is incumbent on the user to ensure the suitability of the
actual procedures used by means of suitable measures (e.g. val-
idation, routine monitoring, check of material compatibility).
5.5 Dismantling
The product must be prepared while it is dismantled.
1. Disconnect the connection cable from the handle (no pic-
ture).
2. Coagulation forceps: Close the handpiece.
3. Press the release button and pull the insert out of the shaft
tube.
3.
5.
4.
3.
2.
OBJ_DOKU-25953-005.fm Seite 21 Dienstag, 28. Juli 2015 1:13 13
Summary of Contents for 20195-080
Page 2: ...OBJ_DOKU 25953 005 fm Seite 2 Dienstag 28 Juli 2015 1 13 13...
Page 26: ...26 OBJ_DOKU 25953 005 fm Seite 26 Dienstag 28 Juli 2015 1 13 13...
Page 77: ...77 Erbe 1 2 800 Vp 3 OBJ_DOKU 25953 005 fm Seite 77 Dienstag 28 Juli 2015 1 13 13...
Page 84: ...84 6 7 1 8 9 1 2 3 4 5 6 OBJ_DOKU 25953 005 fm Seite 84 Dienstag 28 Juli 2015 1 13 13...
Page 110: ...110 OBJ_DOKU 25953 005 fm Seite 110 Dienstag 28 Juli 2015 1 13 13...
Page 122: ...122 OBJ_DOKU 25953 005 fm Seite 122 Dienstag 28 Juli 2015 1 13 13...
Page 134: ...134 OBJ_DOKU 25953 005 fm Seite 134 Dienstag 28 Juli 2015 1 13 13...
Page 158: ...158 OBJ_DOKU 25953 005 fm Seite 158 Dienstag 28 Juli 2015 1 13 13...
Page 170: ...170 OBJ_DOKU 25953 005 fm Seite 170 Dienstag 28 Juli 2015 1 13 13...
Page 173: ...173 Erbe 1 2 800 Vp 3 OBJ_DOKU 25953 005 fm Seite 173 Dienstag 28 Juli 2015 1 13 13...
Page 185: ...185 Erbe 1 2 800 Vp 3 OBJ_DOKU 25953 005 fm Seite 185 Dienstag 28 Juli 2015 1 13 13...
Page 193: ...OBJ_DOKU 25953 005 fm Seite 193 Dienstag 28 Juli 2015 1 13 13...